Med J Malaysia, 2023 Dec;78(7):901-906.
PMID: 38159926

Abstract

INTRODUCTION: The sensorineural acuity level (SAL) test was developed as an alternative assessment to estimate bone conduction (BC) thresholds in cases where masking problems occur in pure tone audiometry (PTA). Nevertheless, prior to its clinical application, the respective SAL normative data must be made available. As such, the present study was carried out to establish SAL normative data using an insert earphone and two different commercially available bone transducers. Additionally, to determine the effect of earphone type on SAL test results, it was also of interest to compare the present study's findings with those of a previous study (that used a headphone to derive SAL normative data).

MATERIALS AND METHODS: In this repeated-measures study, 40 Malaysian adults (aged 19-26 years) with normal hearing bilaterally (based on PTA results) were enrolled. They then underwent the SAL test based on the recommended protocol by Jerger and Tillman (1960). The SAL normative data for each ear were obtained by calculating the differences between air conduction (AC) thresholds in quiet and AC thresholds in noise by means of insert earphone, B71 and B81 bone vibrators.

RESULTS: The SAL normative values were comparable between the ears (p > 0.05), and the data were pooled for subsequent analyses (n = 80 ears). Relative to B81 bone transducer, B71 bone vibrator produced statistically higher SAL normative data at all frequencies (p < 0.05). The SAL normative values established by the present study were statistically lower than those of the previous study (that utilised headphones) at most of frequencies tested (p < 0.05).

CONCLUSIONS: The SAL normative data produced by the two bone vibrators were significantly different. The SAL normative values were also affected by the type of earphone used. While conducting the SAL test on Malaysian patients, the information provided by this study can be useful to guide the respective clinicians in choosing the appropriate normative data.