Affiliations 

  • 1 Audiology and Speech Pathology Programme, School of Health Sciences, Universiti Sains Malaysia, Kelantan, Malaysia
Ear Hear, 2018 11 22;40(4):1039-1042.
PMID: 30461445 DOI: 10.1097/AUD.0000000000000676

Abstract

OBJECTIVES: Post-auricular muscle response (PAMR) is a large myogenic potential that can be useful in estimating behavioral hearing thresholds when the recording protocol is optimal. The main aim of the present study was to determine the influence of stimulus repetition rate on PAMR threshold.

DESIGN: In this repeated-measures study, 20 normally hearing adults aged between 18 and 30 years were recruited. Tone bursts (500, 1000, 2000, and 4000 Hz) were used to record PAMR thresholds at 3 different stimulus repetition rates (6.1/s, 11.1/s, and 17.1/s).

RESULTS: Statistically higher PAMR thresholds were found for the faster stimulus rate (17.1/s) compared with the slower stimulus rate (6.1/s) (p < 0.05). For all stimulus rates and frequencies, significant correlations were found between PAMR and pure-tone audiometry thresholds (r = 0.62 to 0.82).

CONCLUSIONS: Even though the stimulus rate effect was significant at most of the tested frequencies, the differences in PAMR thresholds between the rates were small (<5 dB). Nevertheless, based on the correlation results, we suggest the use of 11.1/s stimulus rate when recording PAMR thresholds.