Assay interference with heterophilic antibodies has been well described in literature. Rheumatoid factor is known to cause similar interference leading to falsely elevated hormone levels when measured by immunometric methods like enzyme-linked immunosorbent assay (ELISA) or multiplex immunoasays (MIA). We report a case of a 60-year-old male patient with a history of rheumatoid arthritis referred to our endocrine clinic for investigation of hypogonadism and was found to have high serum levels of LH, FSH, SHBG, Prolactin, HCG and TSH. We suspected assay interference and further tests were performed. We used Heteroblock tubes and PEG precipitation to eliminate the interference and the hormone levels post treatment were in the normal range. We believe the interference was caused by high serum levels of rheumatoid factor. Although he was treated with thyroxine for 3 years, we believe he may have been treated inappropriately as his Free T4 level was always normal despite high TSH due to assay interference. Our case illustrates the phenomenon of heterophilic antibody interference likely due to high levels of rheumatoid factor. It is essential for clinicians and endocrinologists in particular to be aware of this possibility when making treatment decisions in these groups of patients.
* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.