This study evaluates the ImageJ software as dosimetric tools for analyzing the film dosimeter in high energy photons and electrons. The percentage depth dose of photons of 6 and 10 MV and electrons of 6 and 9 MeV were measured using the Gafchromic EBT2 film dosimeter. The films were scanned and analyzed using the Verisoft software and ImageJ. The beam profiles at nominal photon and electron beam parameters were also evaluated using the two methods. The PDD measured in ImageJ at high energy photons were in good agreement within 1% percentage of discrepancy at all depths in comparison to the Verisoft software. The PDD measured in ImageJ at high energy electrons also showed good agreement to Verisoft software within 8% percentage of discrepancy at all depths. The measured flatness of beam profiles at Dmax, R50, R80 and R90 in ImageJ were also in good agreement to Verisoft software with flatness value between 4 and 8%. The results indicated the suitability of ImageJ software as dosimetric tool for analyzing EBT2 film dosimeter at high energy photon and electrons.