J Med Radiat Sci, 2020 Sep;67(3):170-176.
PMID: 32219989 DOI: 10.1002/jmrs.387

Abstract

INTRODUCTION: 3D-printed imaging phantoms are now increasingly available and used for computed tomography (CT) dose optimisation study and image quality analysis. The aim of this study was to evaluate the integrated 3D-printed cardiac insert phantom when evaluating iterative reconstruction (IR) algorithm in coronary CT angiography (CCTA) protocols.

METHODS: The 3D-printed cardiac insert phantom was positioned into a chest phantom and scanned with a 16-slice CT scanner. Acquisitions were performed with CCTA protocols using 120 kVp at four different tube currents, 300, 200, 100 and 50 mA (protocols A, B, C and D, respectively). The image data sets were reconstructed with a filtered back projection (FBP) and three different IR algorithm strengths. The image quality metrics of image noise, signal-noise ratio (SNR) and contrast-noise ratio (CNR) were calculated for each protocol.

RESULTS: Decrease in dose levels has significantly increased the image noise, compared to FBP of protocol A (P 