Displaying publications 1 - 20 of 28 in total

Abstract:
Sort:
  1. Ghazali L, Mohd Yusof MYP, Norman NH
    J Orthod, 2021 03;48(1):5-12.
    PMID: 33200660 DOI: 10.1177/1465312520971641
    OBJECTIVES: To investigate the effect of reducing scanning parameters of digital dental panoramic tomogram (DPT) and lateral cephalometric (LC) radiographs on quality and diagnostic performance of the images.

    DESIGN: Single-centre prospective two-arm parallel randomised controlled trial.

    SETTING: Orthodontic Clinic, Faculty of Dentistry, Universiti Teknologi MARA, Selangor, Malaysia.

    PARTICIPANTS: Adult orthodontic patients aged 18-35 years, indicated for DPT and LC, who were fit and healthy with a body mass index of 18.5-25.0, not contraindicated to radiographic examination, not pregnant, and did not have a history of facial or skeletal abnormalities or bone diseases were included.

    METHODS: Thirty-eight adult orthodontic patients were randomised into control and intervention groups. DPT and LC radiographs in the control group were obtained using standard scanning parameters as prescribed by the manufacturer using Orthopantomograph® OP300 by Instrumentarium. Scanning parameters in the intervention group were reduced by 60% for DPT (60 kV, 3.2 mA) and 30% for LC (85 kV, 8 mA). A five-point rating scale was used for the assessment of image quality. Images were evaluated for diagnostic performance by detection of anatomical landmarks. Mann-Whitney test was performed to compare the quality and diagnostic performance of the images and the observer agreement was assessed using the intraclass correlation coefficient (ICC).

    RESULTS: For image quality, the control group produced slightly lower median scores (DPT 2.0, LC 2.0) compared to the intervention group (DPT 2.0, LC 3.0). For diagnostic performance, both groups showed similar median scores (DPT 21.0, LC 32.0). The differences between control and intervention groups for both modalities were not statistically significant. The average scores for intra-observer agreement were excellent (ICC 0.917) and inter-observer agreement was good (ICC 0.822).

    CONCLUSION: Minimising radiation exposure by reducing scanning parameters on digital DPT by 60% and LC by 30% on Intsrumentarium 300 OP did not affect the quality and diagnostic performance of the images. Thus, scanning parameters on digital DPT and LC should be reduced when taking radiographs.

    Matched MeSH terms: Radiography, Dental, Digital*
  2. Ahmed N, Abbasi MS, Zuberi F, Qamar W, Halim MSB, Maqsood A, et al.
    Biomed Res Int, 2021;2021:9751564.
    PMID: 34258283 DOI: 10.1155/2021/9751564
    Objective: The objective of this systematic review was to investigate the quality and outcome of studies into artificial intelligence techniques, analysis, and effect in dentistry.

    Materials and Methods: Using the MeSH keywords: artificial intelligence (AI), dentistry, AI in dentistry, neural networks and dentistry, machine learning, AI dental imaging, and AI treatment recommendations and dentistry. Two investigators performed an electronic search in 5 databases: PubMed/MEDLINE (National Library of Medicine), Scopus (Elsevier), ScienceDirect databases (Elsevier), Web of Science (Clarivate Analytics), and the Cochrane Collaboration (Wiley). The English language articles reporting on AI in different dental specialties were screened for eligibility. Thirty-two full-text articles were selected and systematically analyzed according to a predefined inclusion criterion. These articles were analyzed as per a specific research question, and the relevant data based on article general characteristics, study and control groups, assessment methods, outcomes, and quality assessment were extracted.

    Results: The initial search identified 175 articles related to AI in dentistry based on the title and abstracts. The full text of 38 articles was assessed for eligibility to exclude studies not fulfilling the inclusion criteria. Six articles not related to AI in dentistry were excluded. Thirty-two articles were included in the systematic review. It was revealed that AI provides accurate patient management, dental diagnosis, prediction, and decision making. Artificial intelligence appeared as a reliable modality to enhance future implications in the various fields of dentistry, i.e., diagnostic dentistry, patient management, head and neck cancer, restorative dentistry, prosthetic dental sciences, orthodontics, radiology, and periodontics.

    Conclusion: The included studies describe that AI is a reliable tool to make dental care smooth, better, time-saving, and economical for practitioners. AI benefits them in fulfilling patient demand and expectations. The dentists can use AI to ensure quality treatment, better oral health care outcome, and achieve precision. AI can help to predict failures in clinical scenarios and depict reliable solutions. However, AI is increasing the scope of state-of-the-art models in dentistry but is still under development. Further studies are required to assess the clinical performance of AI techniques in dentistry.

    Matched MeSH terms: Radiography, Dental/methods*; Radiography, Dental/trends*
  3. Western JS, Gupta VV, Ramachandra SS
    Compend Contin Educ Dent, 2019 Mar;40(3):172-177.
    PMID: 30829499
    This case report describes the management of a mandibular first molar with an additional distolingual root (radix entomolaris) and grade III cervical enamel projection through a multidisciplinary approach. Diagnosis for the case was endodontic-periodontal lesion due to non-vitality and associated advanced periodontal destruction. The patient was treated with drainage of the periodontal abscess with adjunct antibiotics, phase I periodontal therapy, endodontic therapy, radiculoplasty, regenerative periodontal therapy, replacement of the missing right mandibular second molar, and long-term maintenance. Follow-up of the patient up to 9 months has been uneventful. Cases of advanced periodontal destruction typically show some degree of tooth mobility, which was absent in this case. The article discusses the tripod effect as well as the increased surface area for periodontal attachment provided by the additional root contributing to the non-mobility of the involved tooth.
    Matched MeSH terms: Radiography, Dental
  4. Pratima B, Chandan GD, Nidhi T, Nitish I, Sankriti M, Nagaveni S, et al.
    J Indian Soc Pedod Prev Dent, 2018 9 25;36(3):308-314.
    PMID: 30246755 DOI: 10.4103/JISPPD.JISPPD_1132_17
    Aim: The present study is an attempt to compare and evaluate postoperative assessment of diode laser zinc oxide eugenol (ZOE) pulpotomy and diode laser mineral trioxide aggregate (MTA) pulpotomy procedures in children.

    Materials and Methods: Forty carious primary molars indicated for pulpotomy within the age group of 4-9 years were selected and divided into two groups of 20 each using simple randomization, Group 1: Diode laser MTA and Group 2: Diode laser ZOE pulpotomy. The teeth were evaluated clinically for 1 year at 3, 6, and 12 months interval and radiologically for 6 and 12 months.

    Results: Clinically and radiographically, 100% teeth treated with diode laser MTA and 94% treated with diode laser ZOE were considered successful after 12-month follow-up interval. No significant difference was seen between two groups.

    Conclusion: Despite the success rate, the cost factor of diode laser and MTA could be the limiting factor in its judicious use in pulpotomy procedure.

    Matched MeSH terms: Radiography, Dental
  5. Alyahyawi A, Jupp T, Alkhorayef M, Bradley DA
    Appl Radiat Isot, 2018 Aug;138:45-49.
    PMID: 28780202 DOI: 10.1016/j.apradiso.2017.07.011
    In the modern clinical practice of diagnostic radiology there is a growing demand for radiation dosimetry, it also being recognized that with increasing use of X-ray examinations additional population dose will result, accompanied by an additional albeit low potential for genetic consequences. At the doses typical of diagnostic radiology there is also a low statistical risk for cancer induction; in adhering to best practice, to be also implied is a low but non-negligible potential for deterministic sensitive organ responses, including in regard to the skin and eyes. Risk reduction is important, in line with the principle of ALARP, both in regard to staff and patients alike; for the latter modern practice is usually guided by Dose Reference Levels (DRL) while for the former and members of the public, legislated controls (supported by safe working practices) pertain. As such, effective, reliable and accurate means of dosimetry are required in support of these actions. Recent studies have shown that Ge-doped-silica glass fibres offer several advantages over the well-established phosphor-based TL dosimeters (TLD), including excellent sensitivity at diagnostic doses as demonstrated herein, low fading, good reproducibility and re-usability, as well as representing a water impervious, robust dosimetric system. In addition, these silica-based fibres show good linearity over a wide dynamic range of dose and dose-rate and are directionally independent. In the present study, we investigate tailor made doped-silica glass thermoluminescence (TL) for applications in medical diagnostic imaging dosimetry. The aim is to develop a dosimeter of sensitivity greater than that of the commonly used LiF (Mg,Ti) phosphor. We examine the ability of such doped glass media to detect the typically low levels of radiation in diagnostic applications (from fractions of a mGy through to several mGy or more), including, mammography and dental radiology, use being made of x-ray tubes located at the Royal Surrey County Hospital. We further examine dose-linearity, energy response and fading.
    Matched MeSH terms: Radiography, Dental
  6. Nawi N, Mohamed AM, Marizan Nor M, Ashar NA
    J Orofac Orthop, 2018 Jan;79(1):19-27.
    PMID: 29116344 DOI: 10.1007/s00056-017-0111-3
    OBJECTIVE: The aim of the present study was to determine the overall reliability and validity of arch parameters measured digitally compared to conventional measurement.

    METHODS: A sample of 111 plaster study models of Down syndrome (DS) patients were digitized using a blue light three-dimensional (3D) scanner. Digital and manual measurements of defined parameters were performed using Geomagic analysis software (Geomagic Studio 2014 software, 3D Systems, Rock Hill, SC, USA) on digital models and with a digital calliper (Tuten, Germany) on plaster study models. Both measurements were repeated twice to validate the intraexaminer reliability based on intraclass correlation coefficients (ICCs) using the independent t test and Pearson's correlation, respectively. The Bland-Altman method of analysis was used to evaluate the agreement of the measurement between the digital and plaster models.

    RESULTS: No statistically significant differences (p > 0.05) were found between the manual and digital methods when measuring the arch width, arch length, and space analysis. In addition, all parameters showed a significant correlation coefficient (r ≥ 0.972; p 

    Matched MeSH terms: Radiography, Dental, Digital*
  7. Qabbani AA, Razak NHA, Kawas SA, Sheikh Abdul Hamid S, Wahbi S, Samsudin AR
    J Craniofac Surg, 2017 Jun;28(4):e318-e325.
    PMID: 28230596 DOI: 10.1097/SCS.0000000000003569
    The aim of this study was to determine the efficacy of immediate implant placement with alveolar bone augmentation on socket preservation following atraumatic tooth extraction and comparing it with a tooth alveolar socket that was allowed to heal in a conventional way.Twenty medically fit patients (8 males and 12 females aged between 18 and 40 years) who needed noncomplicated tooth extraction of mandibular premolar teeth were divided randomly and equally into 2 groups. In Group I, the empty extraction socket was left untreated and allowed to heal in a conventional way. In Group II, the immediate implant was placed and the gap between the implant and the inner buccal plate surface of the socket wall was filled with lyophilized bovine bone granules and the wound was covered with pericardium membrane. The patients were followed up clinically and radiologically for regular reviews at 1 week, 3 months, and 9 months postoperative. Cone beam computerized tomography images of the alveolar ridge and socket were analyzed to determine the structural changes of the alveolar ridge. Resonance frequency analysis was measured at 9 months for Group II to assess the degree of secondary stability of the implants by using Osstell machine.A significant difference of bone resorption of 1.49 mm (confidence interval, CI 95%, 0.63-2.35) was observed within the control group at 3 months, and 1.84 mm (P ≤ 0.05) at 9 months intervals. No significant changes of bone resorption were observed in Group II. Comparison between groups showed a highly significant difference at 3 months; 2.56 mm (CI 95% 4.22-0.90) and at 9 months intervals; 3.2 mm (CI 95%, 4.70-1.62) P ≤ 0.001 between Group I and II. High resonance frequency analysis values were observed at 9 months postoperative in Group II.In conclusion, the insertion of immediate implants in fresh extraction sockets together with grafting the circumferential gap between the bony socket wall and the implant surface with bovine bone granules was able to preserve a greater amount of alveolar ridge volume when compared with an extraction socket that was left to heal in a conventional way.
    Matched MeSH terms: Radiography, Dental
  8. Varghese E, Samson RS, Kumbargere SN, Pothen M
    BMJ Case Rep, 2017 May 22;2017.
    PMID: 28536237 DOI: 10.1136/bcr-2017-220506
    Matched MeSH terms: Radiography, Dental/methods*
  9. Kanagasingam S, Hussaini HM, Soo I, Baharin S, Ashar A, Patel S
    Int Endod J, 2017 May;50(5):427-436.
    PMID: 27063356 DOI: 10.1111/iej.12651
    AIM: To compare the accuracy of film and digital periapical radiography (PR) in detecting apical periodontitis (AP) using histopathological findings as a reference standard.

    METHODOLOGY: Jaw sections containing 67 teeth (86 roots) were collected from nine fresh, unclaimed bodies that were due for cremation. Imaging was carried out to detect AP lesions using film and digital PR with a centred view (FP and DP groups); film and digital PR combining central with 10˚ mesially and distally angled (parallax) views (FPS and DPS groups). All specimens underwent histopathological examination to confirm the diagnosis of AP. Sensitivity, specificity and predictive values of PR were analysed using rater mean (n = 5). Receiver operating characteristics (ROC) analysis was carried out.

    RESULTS: Sensitivity was 0.16, 0.37, 0.27 and 0.38 for FP, FPS, DP and DPS, respectively. Both FP and FPS had specificity and positive predictive values of 1.0, whilst DP and DPS had specificity and positive predictive values of 0.99. Negative predictive value was 0.36, 0.43, 0.39 and 0.44 for FP, FPS, DP and DPS, respectively. Area under the curve (AUC) for the various imaging methods was 0.562 (FP), 0.629 (DP), 0.685 (FPS), 0.6880 (DPS).

    CONCLUSIONS: The diagnostic accuracy of single digital periapical radiography was significantly better than single film periapical radiography. The inclusion of two additional horizontal (parallax) angulated periapical radiograph images (mesial and distal horizontal angulations) significantly improved detection of apical periodontitis.

    Matched MeSH terms: Radiography, Dental/methods*; Radiography, Dental, Digital/methods*
  10. Kanagasingam S, Lim CX, Yong CP, Mannocci F, Patel S
    Int Endod J, 2017 May;50(5):417-426.
    PMID: 27063209 DOI: 10.1111/iej.12650
    AIM: To compare the diagnostic accuracy of periapical radiography (PR) and cone beam computed tomography (CBCT) in detecting apical periodontitis (AP) using histopathological findings as a reference standard.

    METHODOLOGY: Jaw sections containing 67 teeth (86 roots) were collected from unclaimed bodies due for cremation. Imaging was carried out to detect AP by digital PR with a central view (DP group), digital PR combining central with 10˚ mesially and distally angled (parallax) views (DPS group) and CBCT scans. All specimens underwent histopathological examination to confirm the diagnosis of AP. Sensitivity, specificity and predictive values of PR and CBCT were analysed using rater mean (n = 5). Receiver-operating characteristic (ROC) analysis was carried out.

    RESULTS: Sensitivity was 0.27, 0.38 and 0.89 for DP, DPS and CBCT scans, respectively. CBCT had specificity and positive predictive value of 1.0 whilst DP and DPS had specificity and positive predictive value of 0.99. The negative predictive value was 0.39, 0.44 and 0.81 for DP, DPS and CBCT scans, respectively. Area under the curve (AUC) for the various imaging methods was 0.629 (DP), 0.688 (DPS), and 0.943 (CBCT).

    CONCLUSIONS: All imaging techniques had similar specificity and positive predictive values. Additional parallax views increased the diagnostic accuracy of PR. CBCT had significantly higher diagnostic accuracy in detecting AP compared to PR, using human histopathological findings as a reference standard.

    Matched MeSH terms: Radiography, Dental, Digital/methods*
  11. Mosleh MA, Baba MS, Malek S, Almaktari RA
    BMC Bioinformatics, 2016 Dec 22;17(Suppl 19):499.
    PMID: 28155649 DOI: 10.1186/s12859-016-1370-5
    BACKGROUND: Cephalometric analysis and measurements of skull parameters using X-Ray images plays an important role in predicating and monitoring orthodontic treatment. Manual analysis and measurements of cephalometric is considered tedious, time consuming, and subjected to human errors. Several cephalometric systems have been developed to automate the cephalometric procedure; however, no clear insights have been reported about reliability, performance, and usability of those systems. This study utilizes some techniques to evaluate reliability, performance, and usability metric using SUS methods of the developed cephalometric system which has not been reported in previous studies.

    METHODS: In this study a novel system named Ceph-X is developed to computerize the manual tasks of orthodontics during cephalometric measurements. Ceph-X is developed by using image processing techniques with three main models: enhancements X-ray image model, locating landmark model, and computation model. Ceph-X was then evaluated by using X-ray images of 30 subjects (male and female) obtained from University of Malaya hospital. Three orthodontics specialists were involved in the evaluation of accuracy to avoid intra examiner error, and performance for Ceph-X, and 20 orthodontics specialists were involved in the evaluation of the usability, and user satisfaction for Ceph-X by using the SUS approach.

    RESULTS: Statistical analysis for the comparison between the manual and automatic cephalometric approaches showed that Ceph-X achieved a great accuracy approximately 96.6%, with an acceptable errors variation approximately less than 0.5 mm, and 1°. Results showed that Ceph-X increased the specialist performance, and minimized the processing time to obtain cephalometric measurements of human skull. Furthermore, SUS analysis approach showed that Ceph-X has an excellent usability user's feedback.

    CONCLUSIONS: The Ceph-X has proved its reliability, performance, and usability to be used by orthodontists for the analysis, diagnosis, and treatment of cephalometric.

    Matched MeSH terms: Radiography, Dental, Digital/methods*
  12. Jayaraman J, Wong HM, King NM, Roberts GJ
    J Forensic Leg Med, 2016 Oct;43:26-33.
    PMID: 27441983 DOI: 10.1016/j.jflm.2016.07.007
    BACKGROUND: Many countries have recently experienced a rapid increase in the demand for forensic age estimates of unaccompanied minors. Hong Kong is a major tourist and business center where there has been an increase in the number of people intercepted with false travel documents. An accurate estimation of age is only possible when a dataset for age estimation that has been derived from the corresponding ethnic population. Thus, the aim of this study was to develop and validate a Reference Data Set (RDS) for dental age estimation for southern Chinese.

    MATERIALS AND METHODS: A total of 2306 subjects were selected from the patient archives of a large dental hospital and the chronological age for each subject was recorded. This age was assigned to each specific stage of dental development for each tooth to create a RDS. To validate this RDS, a further 484 subjects were randomly chosen from the patient archives and their dental age was assessed based on the scores from the RDS. Dental age was estimated using meta-analysis command corresponding to random effects statistical model. Chronological age (CA) and Dental Age (DA) were compared using the paired t-test.

    RESULTS: The overall difference between the chronological and dental age (CA-DA) was 0.05 years (2.6 weeks) for males and 0.03 years (1.6 weeks) for females. The paired t-test indicated that there was no statistically significant difference between the chronological and dental age (p > 0.05).

    CONCLUSION: The validated southern Chinese reference dataset based on dental maturation accurately estimated the chronological age.

    Matched MeSH terms: Radiography, Dental, Digital
  13. Tai MW, Chong ZF, Asif MK, Rahmat RA, Nambiar P
    Leg Med (Tokyo), 2016 Sep;22:42-8.
    PMID: 27591538 DOI: 10.1016/j.legalmed.2016.07.009
    This study was to compare the suitability and precision of xerographic and computer-assisted methods for bite mark investigations. Eleven subjects were asked to bite on their forearm and the bite marks were photographically recorded. Alginate impressions of the subjects' dentition were taken and their casts were made using dental stone. The overlays generated by xerographic method were obtained by photocopying the subjects' casts and the incisal edge outlines were then transferred on a transparent sheet. The bite mark images were imported into Adobe Photoshop® software and printed to life-size. The bite mark analyses using xerographically generated overlays were done by comparing an overlay to the corresponding printed bite mark images manually. In computer-assisted method, the subjects' casts were scanned into Adobe Photoshop®. The bite mark analyses using computer-assisted overlay generation were done by matching an overlay and the corresponding bite mark images digitally using Adobe Photoshop®. Another comparison method was superimposing the cast images with corresponding bite mark images employing the Adobe Photoshop® CS6 and GIF-Animator©. A score with a range of 0-3 was given during analysis to each precision-determining criterion and the score was increased with better matching. The Kruskal Wallis H test showed significant difference between the three sets of data (H=18.761, p<0.05). In conclusion, bite mark analysis using the computer-assisted animated-superimposition method was the most accurate, followed by the computer-assisted overlay generation and lastly the xerographic method. The superior precision contributed by digital method is discernible despite the human skin being a poor recording medium of bite marks.
    Matched MeSH terms: Radiography, Dental
  14. Ahmed HM, Hashem AA
    Int Endod J, 2016 Aug;49(8):724-36.
    PMID: 26174943 DOI: 10.1111/iej.12508
    Anterior teeth may have aberrant anatomical variations in the number of roots and root canals. A review of the literature was conducted using appropriate key words in major endodontic journals to identify the available reported cases as well as experimental and clinical investigations on accessory roots and root canals in anterior teeth. After retrieving the full text of related articles, cross-citations were identified, and the pooled data were then discussed. Results revealed a higher prevalence in accessory root/root canal variations in mandibular anterior teeth than in maxillary counterparts. However, maxillary incisor teeth revealed the highest tendency for accessory root/root canal aberrations caused by anomalies such as dens invaginatus and palato-gingival groove. Primary anterior teeth may also exhibit external and internal anatomical variations in the root, especially maxillary canines. Therefore, dental practitioners should thoroughly assess all teeth scheduled for root canal treatment to prevent the undesirable consequences caused by inadequate debridement of accessory configurations of the root canal system.
    Matched MeSH terms: Radiography, Dental
  15. Kumaresan R, Cugati N, Chandrasekaran B, Karthikeyan P
    J Investig Clin Dent, 2016 Feb;7(1):102-9.
    PMID: 25048008 DOI: 10.1111/jicd.12116
    OBJECTIVE: To evaluate the reliability and validity of Demirjian's, Willems, Nolla's, Haavikko's, and Cameriere's radiographic methods of dental-age estimation in a population of Malaysian children.

    METHODS: A total of 426 dental panoramic radiographs of 5-15-year-old Malaysian children were included in the study. The mean age error and absolute age error for all the methods were calculated and their usability analyzed.

    RESULTS: The Nolla, Willems. and Demirjian methods overestimated the dental age with a mean of 0.97, 0.54, and 0.54 years, respectively, while the Cameriere and Haavikko methods underestimated by 0.41 and 1.31 years, respectively. The Cameriere method was highly precise and accurate in the population of Malaysian children, whereas the Haavikko and Demirjian methods were the least precise and accurate.

    CONCLUSIONS: The Cameriere method of dental-age estimation is highly valid and reliable for Malaysian population, followed by the Willems and Nolla methods.

    Matched MeSH terms: Radiography, Dental, Digital
  16. Lim MY, Lim WW, Rajan S, Nambiar P, Ngeow WC
    Eur Arch Paediatr Dent, 2015 Oct;16(5):397-407.
    PMID: 25894248 DOI: 10.1007/s40368-015-0184-x
    To observe age-related changes in the locations of mandibular foramen (MF) and mental foramen (mf) using dental panoramic tomographs (DPTs).
    Matched MeSH terms: Radiography, Dental, Digital/methods
  17. F AT, Mohammed F
    J Clin Diagn Res, 2014 Jul;8(7):ZD03-4.
    PMID: 25177646 DOI: 10.7860/JCDR/2014/8671.4546
    Susuk or charm needles are a facial cum body art widely practiced among women of Malaysia, Brunei, Singapore, Thailand and Indonesia. These are small, needle-shaped metallic talismans inserted subcutaneously in different parts of the body. The concealed art of susuk was "exposed" by routine radiographic examination in the oral and maxillofacial region. This paper reports two such cases of unusual incidental radiographic finding in dental radiographs which were taken on a routine basis as part of the diagnostic work up. This article will also primarily enlighten the importance of radiographs in detecting such charm needles as the wearer keeps the body art a "hidden secret" thereby avoiding misdiagnosis.
    Matched MeSH terms: Radiography, Dental
  18. Surendran S, Thomas E
    Am J Orthod Dentofacial Orthop, 2014 Jan;145(1):7-14.
    PMID: 24373650 DOI: 10.1016/j.ajodo.2013.09.007
    The objective of this study was to determine whether dental calcification can be used as a first-level diagnostic tool for assessment of skeletal maturity.
    Matched MeSH terms: Radiography, Dental, Digital/methods
  19. Al-Juboori MJ, AbdulRahaman SB, Hassan A
    Implant Dent, 2013 Aug;22(4):351-5.
    PMID: 23811720 DOI: 10.1097/ID.0b013e318296583d
    To detect the correlation between crestal bone resorption and implant stability during healing period using resonance frequency analysis (RFA).
    Matched MeSH terms: Radiography, Dental, Digital/methods
  20. Sabarudin A, Tiau YJ
    Quant Imaging Med Surg, 2013 Feb;3(1):43-8.
    PMID: 23483085 DOI: 10.3978/j.issn.2223-4292.2013.02.07
    This study is designed to compare and evaluate the diagnostic image quality of dental panoramic radiography between conventional and digital systems. Fifty-four panoramic images were collected and divided into three groups consisting of conventional, digital with and without post processing image. Each image was printed out and scored subjectively by two experienced dentists who were blinded to the exposure parameters and system protocols. The evaluation covers of anatomical coverage and structures, density and image contrast. The overall image quality score revealed that digital panoramic with post-processing scored the highest of 3.45±0.19, followed by digital panoramic system without post-processing and conventional panoramic system with corresponding scores of 3.33±0.33 and 2.06±0.40. In conclusion, images produced by digital panoramic system are better in diagnostic image quality than that from conventional panoramic system. Digital post-processing visualization can improve diagnostic quality significantly in terms of radiographic density and contrast.
    Matched MeSH terms: Radiography, Dental, Digital