OBJECTIVE: The purpose of this study was to identify predictors of A4 amplitude and high AVS.
METHODS: We analyzed 64 patients enrolled in MARVEL 2 who had visible P waves on electrocardiogram for assessing A4 amplitude and 40 patients with third-degree AV block for assessing AVS at rest. High AVS was defined as >90% correct atrial-triggered ventricular pacing. The association between clinical factors and echocardiographic parameters with A4 amplitude was investigated using a multivariable model with lasso variable selection. Variables associated with A4 amplitude together with premature ventricular contraction burden, sinus rate, and sinus rate variability (standard deviation of successive differences of P-P intervals [SDSD]) were assessed for association with AVS.
RESULTS: In univariate analysis, low A4 amplitude was inversely related to atrial function assessed by E/A ratio and e'/a' ratio, and was directly related to atrial contraction excursion (ACE) and atrial strain (Ɛa) on echocardiography (all P ≤.05). The multivariable lasso regression model found coronary artery bypass graft history, E/A ratio, ACE, and Ɛa were associated with low A4 amplitude. E/A ratio and SDSD were multivariable predictors of high AVS, with >90% probability if E/A <0.94 and SDSD <5 bpm.
CONCLUSION: Clinical parameters and echocardiographic markers of atrial function are associated with A4 signal amplitude. High AVS can be predicted by E/A ratio <0.94 and low sinus rate variability at rest.
OBJECTIVE: The purpose of this study was to evaluate the system's performance during treadmill tests to maximum exertion in a subset of patients within the Micra Transcatheter Pacing Study.
METHODS: Patients underwent treadmill testing at 3 or 6 months postimplant with algorithm programming at physician discretion. Normalized sensor rate (SenR) relative to the programmed upper sensor rate was modeled as a function of normalized workload in metabolic equivalents (METS) relative to maximum METS achieved during the test. A normalized METS and SenR were determined at the end of each 1-minute treadmill stage. The proportionality of SenR to workload was evaluated by comparing the slope from this relationship to the prospectively defined tolerance margin (0.65-1.35).
RESULTS: A total of 69 treadmill tests were attempted by 42 patients at 3 and 6 months postimplant. Thirty tests from 20 patients who completed ≥4 stages with an average slope of 0.86 (90% confidence interval 0.77-0.96) confirmed proportionality to workload. On an individual test basis, 25 of 30 point estimates (83.3%) had a normalized slope within the defined tolerance range (range 0.46-1.08).
CONCLUSION: Accelerometer-based rate adaptive pacing was proportional to workload, thus confirming rate adaptive pacing commensurate to workload is achievable with an entirely intracardiac pacing system.
OBJECTIVE: The purpose of this study was to sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing.
METHODS: The Micra Accelerometer Sensor Sub-Study (MASS) and MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. The Micra Atrial TRacking Using A Ventricular AccELerometer (MARVEL) study was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. Atrioventricular synchrony (AVS) was measured during 30 minutes of rest and during VVI pacing. AVS was defined as a P wave visible on surface ECG followed by a ventricular event <300 ms.
RESULTS: A total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range 0-41.4). High-degree AV block was present in 33 patients, whereas 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (95% confidence interval 81.8%-90.9%), 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P
CASE PRESENTATION: A 59-year-old man presented with the complaint of recurrent dizziness associated with meals. A 24-h ambulatory ECG recording confirmed an episode of p-wave asystole at the time of food intake. Oesophagogastroduodenoscopy with balloon inflation in the mid to lower oesophagus resulted in a 5.6 s sinus pause. The patient's symptoms resolved completely following insertion of a permanent dual chamber pacemaker.
CONCLUSIONS: Swallow syncope is extremely rare, but still needs to be considered during diagnostic workup. It is commonly associated with gastro-intestinal disease. Permanent pacemaker implantation is the first line treatment.
OBJECTIVE: To describe the technical difficulties on performing posterior spinal fusion (PSF) on a pacemaker-dependent patient with complete congenital heart block and right thoracic scoliosis.
SUMMARY OF BACKGROUND DATA: Congenital complete heart block requires pacemaker implantation at birth through thoracotomy, which can result in scoliosis. Corrective surgery in this patient was challenging. Height gain after corrective surgery may potentially cause lead dislodgement. The usage of monopolar electrocautery may interfere with the function of the implanted cardiac device.
METHODS: A 17-year-old boy was referred to our institution for the treatment of right thoracic scoliosis of 70°. He had underlying complete congenital heart block secondary to maternal systemic lupus erythematosus. Pacemaker was implanted through thoracotomy since birth and later changed for four times. PSF was performed by two attending surgeons with a temporary pacing inserted before the surgery. The monopolar electrocautery device was used throughout the surgery.
RESULTS: The PSF was successfully performed without any technical issues and complications. Postoperatively, his permanent pacemaker was functioning normally. Three days later, he was recovering well and was discharged home from hospital.
CONCLUSION: This case indicates that PSF can be performed successfully with thoughtful anticipation of technical difficulties on a pacemaker-dependent patient with underlying congenital heart block.
LEVEL OF EVIDENCE: 5.
CASE PRESENTATION: A 37 year old Malaysian Chinese parturient was admitted at 25 weeks gestation following a scan which suggested intrauterine death and placenta accreta. She was diagnosed to have congenital complete heart block after her first delivery eight years previously but a pacemaker was never inserted. These medical conditions make her extremely likely to experience massive bleeding and haemodynamic instability. Among the measures taken to optimise her pre-operatively were the insertion of a temporary intravenous pacemaker and embolization of the uterine arteries to minimize peri-operative blood loss. She successfully underwent surgery under general anesthesia, which was relatively uneventful and was discharged well on the fourth post-operative day.
CONCLUSION: Congenital heart block in pregnancies in the presence of potential massive bleeding is best managed by a team, with meticulous pre-operative optimization. Suggested strategies would include insertion of a temporary pacemaker and embolization of the uterine arteries to reduce the risk of the patient getting into life threatening situations.