OBJECTIVE: The purpose of this report was to describe the prespecified long-term safety objective of Micra at 12 months and electrical performance through 24 months.
METHODS: The Micra Transcatheter Pacing Study was a prospective single-arm study designed to assess the safety and efficacy of the Micra VVIR leadless/intracardiac pacemaker. Enrolled patients met class I or II guideline recommendations for de novo ventricular pacing. The long-term safety objective was freedom from a system- or procedure-related major complication at 12 months. A predefined historical control group of 2667 patients with transvenous pacemakers was used to compare major complication rates.
RESULTS: The long-term safety objective was achieved with a freedom from major complication rate of 96.0% at 12 months (95% confidence interval 94.2%-97.2%; P < .0001 vs performance goal). The risk of major complications for patients with Micra (N = 726) was 48% lower than that for patients with transvenous systems through 12 months postimplant (hazard ratio 0.52; 95% confidence interval 0.35-0.77; P = .001). Across subgroups of age, sex, and comorbidities, Micra reduced the risk of major complications compared to transvenous systems. Electrical performance was excellent through 24 months, with a projected battery longevity of 12.1 years.
CONCLUSION: Long-term performance of the Micra transcatheter pacemaker remains consistent with previously reported data. Few patients experienced major complications through 12 months of follow-up, and all patient subgroups benefited as compared to transvenous pacemaker historical control group.
OBJECTIVE: The purpose of this study was to evaluate the system's performance during treadmill tests to maximum exertion in a subset of patients within the Micra Transcatheter Pacing Study.
METHODS: Patients underwent treadmill testing at 3 or 6 months postimplant with algorithm programming at physician discretion. Normalized sensor rate (SenR) relative to the programmed upper sensor rate was modeled as a function of normalized workload in metabolic equivalents (METS) relative to maximum METS achieved during the test. A normalized METS and SenR were determined at the end of each 1-minute treadmill stage. The proportionality of SenR to workload was evaluated by comparing the slope from this relationship to the prospectively defined tolerance margin (0.65-1.35).
RESULTS: A total of 69 treadmill tests were attempted by 42 patients at 3 and 6 months postimplant. Thirty tests from 20 patients who completed ≥4 stages with an average slope of 0.86 (90% confidence interval 0.77-0.96) confirmed proportionality to workload. On an individual test basis, 25 of 30 point estimates (83.3%) had a normalized slope within the defined tolerance range (range 0.46-1.08).
CONCLUSION: Accelerometer-based rate adaptive pacing was proportional to workload, thus confirming rate adaptive pacing commensurate to workload is achievable with an entirely intracardiac pacing system.
BACKGROUND: Leadless pacemakers may be preferred in patients with limited vascular access and high-infection risk, such as patients on hemodialysis.
METHODS: Patients on hemodialysis at the time of Micra implantation attempt (n = 201 of 2,819; 7%) from the Micra Transcatheter Pacing Study investigational device exemption trial, Micra Transcatheter Pacing System Continued Access Study Protocol, and Micra Transcatheter Pacing System Post-Approval Registry were included in the analysis. Baseline characteristics, periprocedural outcomes, and intermediate-term follow-up were summarized.
RESULTS: Patients on hemodialysis at the time of Micra implantation attempt were on average 70.5 ± 13.5 years of age and 59.2% were male. The dialysis patients commonly had hypertension (80%), diabetes (61%), coronary artery disease (39%), and congestive heart failure (27%), and 72% had a condition that the implanting physician felt precluded the use of a transvenous pacemaker. Micra was successfully implanted in 197 patients (98.0%). Reasons for unsuccessful implantation included inadequate thresholds (n = 2) and pericardial effusion (n = 2). The median implantation time was 27 min (interquartile range: 20 to 39 min). There were 3 procedure-related deaths: 1 due to metabolic acidosis following a prolonged procedure duration in a patient undergoing concomitant atrioventricular nodal ablation and 2 deaths occurred in patients who needed surgical repair after perforation. Average follow-up was 6.2 months (range 0 to 26.7 months). No patients had a device-related infection or required device removal because of bacteremia.
CONCLUSIONS: Leadless pacemakers represent an effective pacing option in this challenging patient population on chronic hemodialysis. The risk of infection appears low with an acceptable safety profile. (Micra Transcatheter Pacing Study; NCT02004873; Micra Transcatheter Pacing System Continued Access Study Protocol; NCT02488681; Micra Transcatheter Pacing System Post-Approval Registry; NCT02536118).
OBJECTIVE: The purpose of this study was to describe the natural history of acute elevated Micra vs traditional transvenous lead thresholds.
METHODS: Micra study VVI patients with threshold data (at 0.24 ms) at implant (n = 711) were compared with Capture study patients with de novo transvenous leads at 0.4 ms (n = 538). In both cohorts, high thresholds were defined as >1.0 V and very high as >1.5 V. Change in pacing threshold (0-6 months) with high (1.0 to ≤1.5 V) or very high (>1.5 V) thresholds were compared using the Wilcoxon signed-rank test.
RESULTS: Of the 711 Micra patients, 83 (11.7%) had an implant threshold of >1.0 V at 0.24 ms. Of the 538 Capture patients, 50 (9.3%) had an implant threshold of >1.0 V at 0.40 ms. There were no significant differences in patient characteristics between those with and without an implant threshold of >1.0 V, with the exception of left ventricular ejection fraction in the Capture cohort (high vs low thresholds, 53% vs 58%; P = .011). Patients with an implant threshold of >1.0 V decreased significantly (P < .001) in both cohorts. Micra patients with high and very high thresholds decreased significantly (P < .01) by 1 month, with 87% and 85% having 6-month thresholds lower than the implant value. However, when the capture threshold at implant was >2 V, only 18.2% had a threshold of ≤1 V at 6 months and 45.5% had a capture threshold of >2 V.
CONCLUSIONS: Pacing thresholds in most Micra patients with elevated thresholds decrease after implant. Micra device repositioning may not be necessary if the pacing threshold is ≤2 V.
METHODS AND RESULTS: The primary outcome (risk of cardiac failure, pacemaker syndrome, or syncope related to the Micra system or procedure) was compared between successfully implanted patients from the Micra IDE trial with a primary pacing indication associated with AF or history of AF (AF group) and those without (non-AF group). Among 720 patients successfully implanted with Micra, 228 (31.7%) were in the non-AF group. Reasons for selecting VVI pacing in non-AF patients included an expectation for infrequent pacing (66.2%) and advanced age (27.2%). More patients in the non-AF group had a condition that precluded the use of a transvenous pacemaker (9.6% vs. 4.7%, P = 0.013). Atrial fibrillation patients programmed to VVI received significantly more ventricular pacing compared to non-AF patients (median 67.8% vs. 12.6%; P
MATERIALS AND METHODS: The sample comprised 2,200 dentists from 21 countries. Three scales - Subjective Happiness Scale (SHS), Satisfaction With Life Scale (SWLS), and Affect Balance Scale (ABS) - were used to measure the subjective responses. Data related to demographic and social characteristics were recorded. Mann-Whitney and Kruskal-Wallis tests were used as appropriate. Scales were correlated, and multiple linear regression analyses were employed to identify the independent determinants of SHS, SWLS and ABS. Data were analysed using the SPSS software program; a value of P <0.05 was considered significant.
RESULTS: The overall mean scores of SHS, SWLS and ABS were 18.53 ± 5.06, 23.06 ± 6.25 and 1.26 ± 2.40, respectively, with significant differences found across countries: dentists working in Croatia, Peru and Serbia recorded the highest scores, unlike dentists practicing in Yemen, Syria, and Iraq, who recorded the lowest scores. There were significant, moderately positive correlations between the various scales: SHS and SWLS: r = 0.535, P r = 0.58, P r = 0.533, P
METHODS AND RESULTS: Using the prospective ASIAN-HF (Asian Sudden Cardiac Death in Heart Failure) registry, 5276 patients with symptomatic HF and reduced ejection fraction (HFrEF) from 11 Asian regions and across 3 income regions (high: Hong Kong, Japan, Korea, Singapore, and Taiwan; middle: China, Malaysia, and Thailand; and low: India, Indonesia, and Philippines) were studied. ICD utilization, clinical characteristics, as well as device perception and knowledge, were assessed at baseline among ICD-eligible patients (EF ≤35% and New York Heart Association Class II-III). Patients were followed for the primary outcome of all-cause mortality. Among 3240 ICD-eligible patients (mean age 58.9±12.9 years, 79.1% men), 389 (12%) were ICD recipients. Utilization varied across Asia (from 1.5% in Indonesia to 52.5% in Japan) with a trend toward greater uptake in regions with government reimbursement for ICDs and lower out-of-pocket healthcare expenditure. ICD (versus non-ICD) recipients were more likely to be older (63±11 versus 58±13 year; P<0.001), have tertiary (versus ≤primary) education (34.9% versus 18.1%; P<0.001) and be residing in a high (versus low) income region (64.5% versus 36.5%; P<0.001). Among 2000 ICD nonrecipients surveyed, 55% were either unaware of the benefits of, or needed more information on, device therapy. ICD implantation reduced risks of all-cause mortality (hazard ratio, 0.71; 95% confidence interval, 0.52-0.97) and sudden cardiac deaths (hazard ratio, 0.33; 95% confidence interval, 0.14-0.79) over a median follow-up of 417 days.
CONCLUSIONS: ICDs reduce mortality risk, yet utilization in Asia is low; with disparity across geographic regions and socioeconomic status. Better patient education and targeted healthcare reforms in extending ICD reimbursement may improve access.
CLINICAL TRIAL REGISTRATION: URL: https://clinicaltrials.gov/ct2/show/NCT01633398. Unique identifier: NCT01633398.
OBJECTIVE: The purpose of this study was to report the worldwide experience with successful retrieval of the Micra TPS.
METHODS: A list of all successful retrievals of the currently available leadless pacemakers (LPs) was obtained from the manufacturer of Micra TPS. Pertinent details of retrieval, such as indication, days postimplantation, equipment used, complications, and postretrieval management, were obtained from the database collected by the manufacturer. Other procedural details were obtained directly from the operators at each participating site.
RESULTS: Data from the manufacturer consisted of 40 successful retrievals of the Micra TPS. Operators for 29 retrievals (73%) provided the consent and procedural details. Of the 29 retrievals, 11 patients underwent retrieval during the initial procedure (immediate retrieval); the other 18 patients underwent retrieval during a separate procedure (delayed retrieval). Median duration before delayed retrieval was 46 days (range 1-95 days). The most common reason for immediate retrieval was elevated pacing threshold after tether removal. The most common reasons for delayed retrieval included elevated pacing threshold at follow-up, endovascular infection, and need for transvenous device. Mean procedure duration was 63.11 ± 56 minutes. All retrievals involved snaring via a Micra TPS delivery catheter or steerable sheath. No serious complications occurred during the reported retrievals.
CONCLUSION: Early retrieval of the Micra TPS is feasible and safe.