Displaying all 18 publications

Abstract:
Sort:
  1. Ng TS
    Med J Malaysia, 1987 Jun;42(2):86-9.
    PMID: 3503195
    Temporary cardiac pacing is an established therapeutic modality in the treatment of heart block and bradyarrythmias due to various causes like drugs and metabolic causes and prior to permanent pacing. Temporary pacing using fluoroscopy is best but the image intensifier is not available to all medical units. The method of temporary pacing at the bedside is described in detail.
    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  2. Garweg C, Khelae SK, Chan JYS, Chinitz L, Ritter P, Johansen JB, et al.
    J Cardiovasc Electrophysiol, 2021 07;32(7):1947-1957.
    PMID: 33928713 DOI: 10.1111/jce.15061
    INTRODUCTION: MARVEL 2 assessed the efficacy of mechanical atrial sensing by a ventricular leadless pacemaker, enabling a VDD pacing mode. The behavior of the enhanced MARVEL 2 algorithm during variable atrio-ventricular conduction (AVC) and/or arrhythmias has not been characterized and is the focus of this study.

    METHODS: Of the 75 patients enrolled in the MARVEL 2 study, 73 had a rhythm assessment and were included in the analysis. The enhanced MARVEL 2 algorithm included a mode-switching algorithm that automatically switches between VDD and ventricular only antibradycardia pacing (VVI)-40 depending upon AVC status.

    RESULTS: Forty-two patients (58%) had persistent third degree AV block (AVB), 18 (25%) had 1:1 AVC, 5 (7%) had variable AVC status, and 8 (11%) had atrial arrhythmias. Among the 42 patients with persistent third degree AVB, the median ventricular pacing (VP) percentage was 99.9% compared to 0.2% among those with 1:1 AVC. As AVC status changed, the algorithm switched to VDD when the ventricular rate dropped less than 40 bpm. During atrial fibrillation (AF) with ventricular response greater than 40 bpm, VVI-40 mode was maintained. No pauses longer than 1500 ms were observed. Frequent ventricular premature beats reduced the percentage of AV synchrony. During AF, the atrial signal was of low amplitude and there was infrequent sensing.

    CONCLUSION: The mode switching algorithm reduced VP in patients with 1:1 AVC and appropriately switched to VDD during AV block. No pacing safety issues were observed during arrhythmias.

    Matched MeSH terms: Cardiac Pacing, Artificial
  3. Ng WH, Ahmad Z, Ong S
    Med J Malaysia, 1979 Mar;33(3):259-63.
    PMID: 522731
    Matched MeSH terms: Cardiac Pacing, Artificial*
  4. Jin C, Dai Q, Li P, Lam P, Cha YM
    J Cardiovasc Electrophysiol, 2023 Sep;34(9):1933-1943.
    PMID: 37548113 DOI: 10.1111/jce.16013
    INTRODUCTION: Left bundle branch area pacing (LBBP) is a novel conduction system pacing method to achieve effective physiological pacing and an alternative to cardiac resynchronization therapy (CRT) with biventricular pacing (BVP) for patients with heart failure with reduced ejection fraction (HFrEF). We conduted this meta-analysis and systemic review to review current data comparing BVP and LBBP in patients with HFrEF and indications for CRT.

    METHODS: We searched PubMed/Medline, Web of Science, and Cochrane Library from the inception of the database to November 2022. All studies that compared LBBP with BVP in patients with HFrEF and indications for CRT were included. Two reviewers performed study selection, data abstraction, and risk of bias assessment. We calculated risk ratios (RRs) with the Mantel-Haenszel method and mean difference (MD) with inverse variance using random effect models. We assessed heterogeneity using the I2 index, with I2  > 50% indicating significant heterogeneity.

    RESULTS: Ten studies (9 observational studies and 1 randomized controlled trial; 616 patients; 15 centers) published between 2020 and 2022 were included. We observed a shorter fluoroscopy time (MD: 9.68, 95% confidence interval [CI]: 4.49-14.87, I2  = 95%, p 

    Matched MeSH terms: Cardiac Pacing, Artificial
  5. Chin K, Singham KT, Masduki A
    Med J Malaysia, 1984 Jun;39(2):139-42.
    PMID: 6513853
    A retrospective study of the indications for temporary transvenous pacing in the University Hospital, Kuala Lumpur, from 1971 to 1979 is reviewed. There were 111 patients. The main indications for temporary transvenous pacing were, namely,complete heart block (57%), sick sinus syndrome (24%), Mobitz type 11 block (5%) and bifascicularblock (3%).
    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  6. Chin K, Yusof K
    Med J Malaysia, 1987 Jun;42(2):93-6.
    PMID: 3503197
    The clinical course of 68 patients who had temporary transvenous pacing for complete heart block in acute myocardial infarction is reviewed. There were 59 male and 9 female patients, of which 32 were Indians, 20 were Malays and 16 Chinese. 48 patients had inferior infarct, 16 patients developed anterior infarct and 4 patients presented with combined anteriorinferior infarction. The overall mortality for complete heart block in this study is 44%. The mortality pattern is 25% for inferior infarct; 87% for anterior infarct and 100% for combined anterior and inferior infarct.
    Matched MeSH terms: Cardiac Pacing, Artificial*
  7. Steinwender C, Khelae SK, Garweg C, Chan JYS, Ritter P, Johansen JB, et al.
    JACC Clin Electrophysiol, 2020 01;6(1):94-106.
    PMID: 31709982 DOI: 10.1016/j.jacep.2019.10.017
    OBJECTIVES: This study reports on the performance of a leadless ventricular pacemaker with automated, enhanced accelerometer-based algorithms that provide atrioventricular (AV) synchronous pacing.

    BACKGROUND: Despite many advantages, leadless pacemakers are currently only capable of single-chamber ventricular pacing.

    METHODS: The prospective MARVEL 2 (Micra Atrial tRacking using a Ventricular accELerometer 2) study assessed the performance of an automated, enhanced accelerometer-based algorithm downloaded to the Micra leadless pacemaker for up to 5 h in patients with AV block. The primary efficacy objective was to demonstrate the superiority of the algorithm to provide AV synchronous (VDD) pacing versus VVI-50 pacing in patients with sinus rhythm and complete AV block. The primary safety objective was to demonstrate that the algorithm did not result in pauses or heart rates of >100 beats/min.

    RESULTS: Overall, 75 patients from 12 centers were enrolled; an accelerometer-based algorithm was downloaded to their leadless pacemakers. Among the 40 patients with sinus rhythm and complete AV block included in the primary efficacy objective analysis, the proportion of patients with ≥70% AV synchrony at rest was significantly greater with VDD pacing than with VVI pacing (95% vs. 0%; p 

    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  8. Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al.
    Heart Rhythm, 2017 05;14(5):685-691.
    PMID: 28111349 DOI: 10.1016/j.hrthm.2017.01.026
    BACKGROUND: Device repositioning during Micra leadless pacemaker implantation may be required to achieve optimal pacing thresholds.

    OBJECTIVE: The purpose of this study was to describe the natural history of acute elevated Micra vs traditional transvenous lead thresholds.

    METHODS: Micra study VVI patients with threshold data (at 0.24 ms) at implant (n = 711) were compared with Capture study patients with de novo transvenous leads at 0.4 ms (n = 538). In both cohorts, high thresholds were defined as >1.0 V and very high as >1.5 V. Change in pacing threshold (0-6 months) with high (1.0 to ≤1.5 V) or very high (>1.5 V) thresholds were compared using the Wilcoxon signed-rank test.

    RESULTS: Of the 711 Micra patients, 83 (11.7%) had an implant threshold of >1.0 V at 0.24 ms. Of the 538 Capture patients, 50 (9.3%) had an implant threshold of >1.0 V at 0.40 ms. There were no significant differences in patient characteristics between those with and without an implant threshold of >1.0 V, with the exception of left ventricular ejection fraction in the Capture cohort (high vs low thresholds, 53% vs 58%; P = .011). Patients with an implant threshold of >1.0 V decreased significantly (P < .001) in both cohorts. Micra patients with high and very high thresholds decreased significantly (P < .01) by 1 month, with 87% and 85% having 6-month thresholds lower than the implant value. However, when the capture threshold at implant was >2 V, only 18.2% had a threshold of ≤1 V at 6 months and 45.5% had a capture threshold of >2 V.

    CONCLUSIONS: Pacing thresholds in most Micra patients with elevated thresholds decrease after implant. Micra device repositioning may not be necessary if the pacing threshold is ≤2 V.

    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  9. Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JM, et al.
    Heart Rhythm, 2017 05;14(5):702-709.
    PMID: 28192207 DOI: 10.1016/j.hrthm.2017.01.035
    BACKGROUND: Early performance of the Micra transcatheter pacemaker from the global clinical trial reported a 99.2% implant success rate, low and stable pacing capture thresholds, and a low (4.0%) rate of major complications up to 6 months.

    OBJECTIVE: The purpose of this report was to describe the prespecified long-term safety objective of Micra at 12 months and electrical performance through 24 months.

    METHODS: The Micra Transcatheter Pacing Study was a prospective single-arm study designed to assess the safety and efficacy of the Micra VVIR leadless/intracardiac pacemaker. Enrolled patients met class I or II guideline recommendations for de novo ventricular pacing. The long-term safety objective was freedom from a system- or procedure-related major complication at 12 months. A predefined historical control group of 2667 patients with transvenous pacemakers was used to compare major complication rates.

    RESULTS: The long-term safety objective was achieved with a freedom from major complication rate of 96.0% at 12 months (95% confidence interval 94.2%-97.2%; P < .0001 vs performance goal). The risk of major complications for patients with Micra (N = 726) was 48% lower than that for patients with transvenous systems through 12 months postimplant (hazard ratio 0.52; 95% confidence interval 0.35-0.77; P = .001). Across subgroups of age, sex, and comorbidities, Micra reduced the risk of major complications compared to transvenous systems. Electrical performance was excellent through 24 months, with a projected battery longevity of 12.1 years.

    CONCLUSION: Long-term performance of the Micra transcatheter pacemaker remains consistent with previously reported data. Few patients experienced major complications through 12 months of follow-up, and all patient subgroups benefited as compared to transvenous pacemaker historical control group.

    Matched MeSH terms: Cardiac Pacing, Artificial/adverse effects*
  10. Piccini JP, Stromberg K, Jackson KP, Kowal RC, Duray GZ, El-Chami MF, et al.
    Europace, 2019 Nov 01;21(11):1686-1693.
    PMID: 31681964 DOI: 10.1093/europace/euz230
    AIMS: Patient selection is a key component of securing optimal patient outcomes with leadless pacing. We sought to describe and compare patient characteristics and outcomes of Micra patients with and without a primary pacing indication associated with atrial fibrillation (AF) in the Micra IDE trial.

    METHODS AND RESULTS: The primary outcome (risk of cardiac failure, pacemaker syndrome, or syncope related to the Micra system or procedure) was compared between successfully implanted patients from the Micra IDE trial with a primary pacing indication associated with AF or history of AF (AF group) and those without (non-AF group). Among 720 patients successfully implanted with Micra, 228 (31.7%) were in the non-AF group. Reasons for selecting VVI pacing in non-AF patients included an expectation for infrequent pacing (66.2%) and advanced age (27.2%). More patients in the non-AF group had a condition that precluded the use of a transvenous pacemaker (9.6% vs. 4.7%, P = 0.013). Atrial fibrillation patients programmed to VVI received significantly more ventricular pacing compared to non-AF patients (median 67.8% vs. 12.6%; P 

    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  11. Fang F, Luo XX, Zhang Q, Azlan H, Razali O, Ma Z, et al.
    Europace, 2015 Oct;17 Suppl 2:ii47-53.
    PMID: 26842115 DOI: 10.1093/europace/euv130
    Biventricular (BiV) pacing was superior to right ventricular apical (RVA) pacing at extended follow-up in the Pacing to Avoid Cardiac Enlargement (PACE) trial. Early pacing-induced systolic dyssynchrony (DYS) might be related to mid-term result. However, it remains unknown whether early pacing-induced DYS can predict long-term reduction of left ventricular (LV) systolic function.
    Matched MeSH terms: Cardiac Pacing, Artificial/adverse effects*
  12. Garweg C, Khelae SK, Steinwender C, Chan JYS, Ritter P, Johansen JB, et al.
    Heart Rhythm, 2020 12;17(12):2037-2045.
    PMID: 32717315 DOI: 10.1016/j.hrthm.2020.07.024
    BACKGROUND: The MARVEL (Micra Atrial TRacking Using a Ventricular AccELerometer) 2 study assessed the efficacy of atrioventricular (AV) synchronous pacing with a Micra leadless pacemaker. Average atrioventricular synchrony (AVS) was 89.2%. Previously, low amplitude of the Micra-sensed atrial signal (A4) was observed to be a factor of low AVS.

    OBJECTIVE: The purpose of this study was to identify predictors of A4 amplitude and high AVS.

    METHODS: We analyzed 64 patients enrolled in MARVEL 2 who had visible P waves on electrocardiogram for assessing A4 amplitude and 40 patients with third-degree AV block for assessing AVS at rest. High AVS was defined as >90% correct atrial-triggered ventricular pacing. The association between clinical factors and echocardiographic parameters with A4 amplitude was investigated using a multivariable model with lasso variable selection. Variables associated with A4 amplitude together with premature ventricular contraction burden, sinus rate, and sinus rate variability (standard deviation of successive differences of P-P intervals [SDSD]) were assessed for association with AVS.

    RESULTS: In univariate analysis, low A4 amplitude was inversely related to atrial function assessed by E/A ratio and e'/a' ratio, and was directly related to atrial contraction excursion (ACE) and atrial strain (Ɛa) on echocardiography (all P ≤.05). The multivariable lasso regression model found coronary artery bypass graft history, E/A ratio, ACE, and Ɛa were associated with low A4 amplitude. E/A ratio and SDSD were multivariable predictors of high AVS, with >90% probability if E/A <0.94 and SDSD <5 bpm.

    CONCLUSION: Clinical parameters and echocardiographic markers of atrial function are associated with A4 signal amplitude. High AVS can be predicted by E/A ratio <0.94 and low sinus rate variability at rest.

    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  13. Lloyd M, Reynolds D, Sheldon T, Stromberg K, Hudnall JH, Demmer WM, et al.
    Heart Rhythm, 2017 02;14(2):200-205.
    PMID: 27871854 DOI: 10.1016/j.hrthm.2016.11.016
    BACKGROUND: The Micra transcatheter pacemaker was designed to have similar functionality to conventional transvenous VVIR pacing systems. It provides rate adaptive pacing using a programmable 3-axis accelerometer designed to detect patient activity in the presence of cardiac motion.

    OBJECTIVE: The purpose of this study was to evaluate the system's performance during treadmill tests to maximum exertion in a subset of patients within the Micra Transcatheter Pacing Study.

    METHODS: Patients underwent treadmill testing at 3 or 6 months postimplant with algorithm programming at physician discretion. Normalized sensor rate (SenR) relative to the programmed upper sensor rate was modeled as a function of normalized workload in metabolic equivalents (METS) relative to maximum METS achieved during the test. A normalized METS and SenR were determined at the end of each 1-minute treadmill stage. The proportionality of SenR to workload was evaluated by comparing the slope from this relationship to the prospectively defined tolerance margin (0.65-1.35).

    RESULTS: A total of 69 treadmill tests were attempted by 42 patients at 3 and 6 months postimplant. Thirty tests from 20 patients who completed ≥4 stages with an average slope of 0.86 (90% confidence interval 0.77-0.96) confirmed proportionality to workload. On an individual test basis, 25 of 30 point estimates (83.3%) had a normalized slope within the defined tolerance range (range 0.46-1.08).

    CONCLUSION: Accelerometer-based rate adaptive pacing was proportional to workload, thus confirming rate adaptive pacing commensurate to workload is achievable with an entirely intracardiac pacing system.

    Matched MeSH terms: Cardiac Pacing, Artificial/methods*
  14. Valli H, Ahmad S, Fraser JA, Jeevaratnam K, Huang CL
    Exp Physiol, 2017 12 01;102(12):1619-1634.
    PMID: 28960529 DOI: 10.1113/EP086589
    NEW FINDINGS: What is the central question of this study? Can we experimentally replicate atrial pro-arrhythmic phenotypes associated with important chronic clinical conditions, including physical inactivity, obesity, diabetes mellitus and metabolic syndrome, compromising mitochondrial function, and clarify their electrophysiological basis? What is the main finding and its importance? Electrocardiographic and intracellular cardiomyocyte recording at progressively incremented pacing rates demonstrated age-dependent atrial arrhythmic phenotypes in Langendorff-perfused murine Pgc1β-/- hearts for the first time. We attributed these to compromised action potential conduction and excitation wavefronts, whilst excluding alterations in recovery properties or temporal electrophysiological instabilities, clarifying these pro-arrhythmic changes in chronic metabolic disease. Atrial arrhythmias, most commonly manifesting as atrial fibrillation, represent a major clinical problem. The incidence of atrial fibrillation increases with both age and conditions associated with energetic dysfunction. Atrial arrhythmic phenotypes were compared in young (12-16 week) and aged (>52 week) wild-type (WT) and peroxisome proliferative activated receptor, gamma, coactivator 1 beta (Ppargc1b)-deficient (Pgc1β-/- ) Langendorff-perfused hearts, previously used to model mitochondrial energetic disorder. Electrophysiological explorations were performed using simultaneous whole-heart ECG and intracellular atrial action potential (AP) recordings. Two stimulation protocols were used: an S1S2 protocol, which imposed extrasystolic stimuli at successively decremented intervals following regular pulse trains; and a regular pacing protocol at successively incremented frequencies. Aged Pgc1β-/- hearts showed greater atrial arrhythmogenicity, presenting as atrial tachycardia and ectopic activity. Maximal rates of AP depolarization (dV/dtmax ) were reduced in Pgc1β-/- hearts. Action potential latencies were increased by the Pgc1β-/- genotype, with an added interactive effect of age. In contrast, AP durations to 90% recovery (APD90 ) were shorter in Pgc1β-/- hearts despite similar atrial effective recovery periods amongst the different groups. These findings accompanied paradoxical decreases in the incidence and duration of alternans in the aged and Pgc1β-/- hearts. Limiting slopes of restitution curves of APD90 against diastolic interval were correspondingly reduced interactively by Pgc1β-/- genotype and age. In contrast, reduced AP wavelengths were associated with Pgc1β-/- genotype, both independently and interacting with age, through the basic cycle lengths explored, with the aged Pgc1β-/- hearts showing the shortest wavelengths. These findings thus implicate AP wavelength in possible mechanisms for the atrial arrhythmic changes reported here.
    Matched MeSH terms: Cardiac Pacing, Artificial*
  15. Siew KSW, Tan MP, Hilmi IN, Loch A
    BMC Cardiovasc Disord, 2019 08 07;19(1):191.
    PMID: 31391000 DOI: 10.1186/s12872-019-1174-4
    BACKGROUND: Swallow or deglutition syncope is an unusual type of neurally-mediated syncope associated with life-threatening bradyarrhythmia and hypotension. It is a difficult condition to diagnose with commonly delayed diagnosis and management. There is lack of review articles that elucidate the basic demographics, clinical characteristics and management of this rare condition. This publication systematically reviews the 101 case reports published since 1793 on swallow syncope.

    CASE PRESENTATION: A 59-year-old man presented with the complaint of recurrent dizziness associated with meals. A 24-h ambulatory ECG recording confirmed an episode of p-wave asystole at the time of food intake. Oesophagogastroduodenoscopy with balloon inflation in the mid to lower oesophagus resulted in a 5.6 s sinus pause. The patient's symptoms resolved completely following insertion of a permanent dual chamber pacemaker.

    CONCLUSIONS: Swallow syncope is extremely rare, but still needs to be considered during diagnostic workup. It is commonly associated with gastro-intestinal disease. Permanent pacemaker implantation is the first line treatment.

    Matched MeSH terms: Cardiac Pacing, Artificial
  16. Yew KL, Anderson S, Farah R, Lim SH
    Asian Cardiovasc Thorac Ann, 2014 Oct;22(8):979-80.
    PMID: 24887840 DOI: 10.1177/0218492313491583
    Central vein stenosis is not uncommon in hemodialysis-dependent patients as a result of mechanical damage to the vessel walls from prior cannulation. It can cause ipsilateral upper limb swelling and pain, resulting in suboptimal hemodialysis. It is unfortunate for bilateral central vein stenosis to develop concomitantly, and rare in the setting of an in-situ pacemaker. This case illustrates the successful ligation of a nondependent left arteriovenous fistula and stenting of the right subclavian vein with functioning ipsilateral arteriovenous fistula, to overcome the problem of symptomatic bilateral upper limb swelling.
    Matched MeSH terms: Cardiac Pacing, Artificial*
  17. Toh HT
    Am J Chin Med, 1994;22(3-4):275-84.
    PMID: 7872239
    Heart mitochondria freshly isolated from ginseng treated rats respired higher at ADP-induced, state 3 respiratory rates and with greater respiratory indices. These mitochondria were less susceptible to experimentally-induced functional impairment. Control heart mitochondria incubated with ginseng extract also showed that ginseng prevented mitochondria from incubation induced deterioration with NAD-linked substrates. Comparison of force of contraction of isolated, perfused and electrically paced hearts showed that deterioration of the force of heart contraction was consistently smaller throughout the experiment in hearts from ginseng treated rats. These results indicated that Panax ginseng was able to delay experimentally induced heart mitochondrial impairment and muscle contraction deterioration.
    Matched MeSH terms: Cardiac Pacing, Artificial
  18. Ahmad S, Valli H, Edling CE, Grace AA, Jeevaratnam K, Huang CL
    Pflugers Arch., 2017 Dec;469(12):1579-1590.
    PMID: 28821956 DOI: 10.1007/s00424-017-2054-3
    A range of chronic clinical conditions accompany cardiomyocyte energetic dysfunction and constitute independent risk factors for cardiac arrhythmia. We investigated pro-arrhythmic and arrhythmic phenotypes in energetically deficient C57BL mice with genetic ablation of the mitochondrial promoter peroxisome proliferator-activated receptor-γ coactivator-1β (Pgc-1β), a known model of ventricular arrhythmia. Pro-arrhythmic and cellular action potential (AP) characteristics were compared in intact Langendorff-perfused hearts from young (12-16 week) and aged (> 52 week), wild-type (WT) and Pgc-1β -/- mice. Simultaneous electrocardiographic and intracellular microelectrode recordings were made through successive trains of 100 regular stimuli at progressively incremented heart rates. Aged Pgc-1β -/- hearts displayed an increased incidence of arrhythmia compared to other groups. Young and aged Pgc-1β -/- hearts showed higher incidences of alternans in both AP activation (maximum AP upshoot velocity (dV/dt)max and latency), recovery (action potential duration (APD90) and resting membrane potential (RMP) characteristics compared to WT hearts. This was particularly apparent at lower pacing frequencies. These findings accompanied reduced (dV/dt)max and increased AP latency values in the Pgc-1β -/- hearts. APs observed prior to termination of the protocol showed lower (dV/dt)max and longer AP latencies, but indistinguishable APD90 and RMPs in arrhythmic compared to those in non-arrhythmic hearts. APD restitution analysis showed that Pgc-1β -/- and WT hearts showed similar limiting gradients. However, Pgc-1β -/- hearts had shortened plateau AP wavelengths, particularly in aged Pgc-1β -/- hearts. Pgc-1β -/- hearts therefore show pro-arrhythmic instabilities attributable to altered AP conduction and activation rather than recovery characteristics.
    Matched MeSH terms: Cardiac Pacing, Artificial
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links