Biomed Phys Eng Express, 2023 May 30;9(4).
PMID: 37216929 DOI: 10.1088/2057-1976/acd785

Abstract

Objective. To develop an algorithm to measure slice thickness running on three types of Catphan phantoms with the ability to adapt to any misalignment and rotation of the phantoms.Method. Images of Catphan 500, 504, and 604 phantoms were examined. In addition, images with various slice thicknesses ranging from 1.5 to 10.0 mm, distance to the iso-center and phantom rotations were also examined. The automatic slice thickness algorithm was carried out by processing only objects within a circle having a diameter of half the diameter of the phantom. A segmentation was performed within an inner circle with dynamic thresholds to produce binary images with wire and bead objects within it. Region properties were used to distinguish wire ramps and bead objects. At each identified wire ramp, the angle was detected using the Hough transform. Profile lines were then placed on each ramp based on the centroid coordinates and detected angles, and the full-width at half maximum (FWHM) was determined for the average profile. The slice thickness was obtained by multiplying the FWHM by the tangent of the ramp angle (23°).Results. Automatic measurements work well and have only a small difference (<0.5 mm) from manual measurements. For slice thickness variation, automatic measurement successfully performs segmentation and correctly locates the profile line on all wire ramps. The results show measured slice thicknesses that are close (<3 mm) to the nominal thickness at thin slices, but slightly deviated for thicker slices. There is a strong correlation (R2= 0.873) between automatic and manual measurements. Testing the algorithm at various distances from the iso-center and phantom rotation angle also produced accurate results.Conclusion. An automated algorithm for measuring slice thickness on three types of Catphan CT phantom images has been developed. The algorithm works well on various thicknesses, distances from the iso-center, and phantom rotations.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.