PMID: 16908224

Abstract

Nicotine is a major addictive compound in cigarette. Its smoke is rapidly and extensively metabolized to several metabolites in human. Cotinine as a major metabolite of nicotine is commonly used as a biomarker to determine active and passive smokers. Cotinine has a longer half-life ( approximately 20 h) compared to nicotine ( approximately 2h). A simple, sensitive, rapid and high throughput GC-MS method was developed for simultaneous quantification of urinary nicotine and cotinine in passive and active smokers. In the sample preparation method, the analytes and internal standard were first basified and followed by liquid-liquid extraction. Upon completion, anhydrous sodium sulphate was added to the solvent mixture to trap moistures. The clear extract obtained was directly injected into GC-MS, operating under selective ion monitoring (SIM) mode. Calibration curves in the range of 0.5-5000 ng/mL of the analytes in urine matrix were established with linear correlation coefficients (r(2)) greater than 0.997. The limit of detection for both nicotine and cotinine were 0.20 ng/mL. The mean recoveries for nicotine and cotinine were 93.0 and 100.4%, respectively. The within- and between-assay accuracies were between 2.1 and 7.9% for nicotine and between 0.7 and 11.1% for cotinine. Within- and between-assay precisions of 3.3-9.5% for nicotine and 3.4-9.8% for cotinine were also achieved. The method can be used in routine assessment and monitoring of active smoking and exposure to environmental tobacco smoke. The applicability of the assay was demonstrated in a small-scale comparison study between smokers and non-smokers.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.