INTRODUCTION: T-piece resuscitator (TPR) has many advantages compared to self-inflating bag (SIB). Early Continuous Positive Airway Pressure (CPAP) during newborn resuscitation (NR) with TPR at delivery can reduce intubation rate.
METHODS: We speculated that the intubation rate at delivery room was high because SIB had always been used during NR and this can be improved with TPR. Intubation rate of newborn <24 hours of life was deemed high if >50%. An audit was carried out in June 2010 to verify this problem using a check sheet.
RESULTS: 25 neonates without major congenital anomalies who required NR with SIB at delivery were included. Intubation rate of babies <24 hours of life when SIB was used was 68%. Post-intervention audit (August to November 2010) on 25 newborns showed that the intubation rate within 24 hours dropped to 8% when TPR was used. Proportion of intubated babies reduced from 48.3% (2008-2009) to 35.1% (2011-2012), odds ratio 0.58 (95% CI 0.49-0.68). Proportion of neonates on CPAP increased from 63.5% (2008-2009) to 81.0% (2011-2012), odds ratio 2.44 (95% CI 2.03-2.93). Mean ventilation days fell to below 4 days after 2010. Since then, all delivery standbys were accompanied by TPR and it was used for all NR regardless of settings. There was decline in intubation rate secondary to early provision of CPAP with TPR during NR. Mean ventilation days, mortality and length of NICU stay were reduced.
CONCLUSION: This practice should be adopted by all hospitals in the country to achieve Millennium Development Goal 4 (2/3 decline of under 5 mortality rate) by 2015.
Study site: Hospital Tuanku Jaafar, Seremban, Negeri Sembilan, Malaysia