Affiliations 

  • 1 International Islamic University Malaysia
MyJurnal

Abstract

Introduction: Application of dynamic hip screw (DHS) implant for the treatment of unstable intertrochanteric fractures continues to raise concern related to risk of lag screw cut-out with or without subsequent damage to the acetabulum. Measurement of tip-apex distances (TAD) has been recommended to guide the optimal placement of lag screw and to predict subsequent risk of screw cut-out. In this study, the value of TAD was evaluated to verify its usefulness.
Methods: This is a retrospective study of 33 consecutive patients with intertrochanteric fracture treated with DHS. Demographic data of the patients were traced from their case notes. Post-operative radiographs were reviewed by focusing on measurement of TAD on anteroposterior and lateral radiographs. Radiographs at one year follow-up were reviewed to depict any fixation-related failure or complication.
Results: Fifty two percent of patients did not achieved the recommended TAD of ≤ 25mm. The mean post-operative TAD was 25.9mm and elderly patients were likely to achieve TAD of ≤ 25mm. The overall complication rate of 6% was attributed to screw cut-out in two cases. The unstable left-sided fracture was identified to be a potential risk for screw cut-out or migration.
Conclusion: TAD is a valuable measurement to guide optimal placement of lag screw during DHS fixation of intertrochanteric fracture.
KEYWORDS: Intertrochanteric fracture, dynamic hip screw, tip-apex distance, screw cut-out