METHODS: Four miniature manometric assemblies, each containing manometric lumina of either 0.4 or 0.5 mm i.d., were evaluated at 100 and 180 cm lengths. The fidelity of miniature manometric luminal recordings were evaluated in vivo during esophageal peristalsis by using a simultaneous comparison with the standard lumina and an intraluminal strain gauge.
RESULTS: During esophageal peristalsis, miniature manometric lumina recorded the peak amplitude of pressure waves, with an accuracy at perfusion rates of 0.04 mL/min (0.4 mm, i.d.) and 0.15 mL/min (0.5 mm, i.d.).
CONCLUSION: Miniature manometric assemblies of lengths that are practical for use in humans are suitable for recording esophageal peristalsis.
Methods: Patients with chronic constipation (Rome III) were evaluated for DD with HRPT and WM during bearing-down "on-bed" without inflated rectal balloon and "on-commode (toilet)" with 60-mL inflated rectal balloon. Eleven healthy volunteers were also evaluated.
Results: Ninety-three of 117 screened participants (F/M = 77/16) were included. Balloon expulsion time was abnormal (> 60 seconds) in 56% (mean 214.4 seconds). A modest correlation between HRPT and WM was observed for sphincter length (R = 0.4) and likewise agreement between dyssynergic subtypes (κ = 0.4). During bearing down, 2 or more anal pressure-segments (distal and proximal) could be appreciated and their expansion measured with HRPT but not WM. In constipated vs healthy participants, the proximal segment was more expanded (2.0 cm vs 1.0 cm, P = 0.003) and of greater pressure (94.8 mmHg vs 54.0 mmHg, P = 0.010) during bearing down on-commode but not on-bed.
Conclusions: Because of its better resolution, HRPT may identify more structural and functional abnormalities including puborectal dysfunction (proximal expansion) than WM. Bearing down on-commode with an inflated rectal balloon may provide additional dimension in characterizing DD.
METHODOLOGY: This was a prospective observational study. Convenience sampling method was used to recruit all HD patients who had definitive pullthrough from January 2019 to December2020 in our institution. High-resolution anorectal manometry (HRAM) was used to record anal resting pressure (ARP), length of high-pressure zone (HPZ), and presence/absence of recto-anal inhibitory reflex (RAIR). The Paediatric Incontinence/Constipation Scoring System (PICSS) was scored for all participants. PICSS is a validated questionnaire with scores mapped to an age-specific normogram to denote constipation, incontinence, and their combinations. Non-parametric and chi-square tests at significance p<0.05 were conducted to examine the relationship between PICSS categories and manometry findings. Ethical approval was obtained.
RESULTS: There were 32 participants (30 boys). Median age at participation was 26.5 months (range: 13.8-156). Twenty-four (75%) had transanal pullthrough, 8(25%) underwent Duhamel procedure. PICSS scored 10(31.3%) as normal, 8(25%) as constipation, 10(31.3%) as incontinent, and 4(12.5%) as mixed. RAIR was present in 12 patients (37.5%). HPZ, maximum ARP, mean ARP were comparable across all PICSS groups without statistically significant differences. Presence of RAIR was not significantly associated with any PICSS groups (p = 0.13).
CONCLUSION: Bowel function does not appear to be significantly associated with HRAM findings after definitive pullthrough for HD, but our study is limited by small sample size. RAIR was present in 37.5% patients after pullthrough.
LEVEL OF EVIDENCE: Level II.
METHODS: Adult patients with persistent gastroesophageal reflux disease (GERD) symptoms undergoing HRM and pH-impedance or wireless pH study off proton pump inhibitor were prospectively studied between July 2021 and March 2022. After the HRM Chicago 4.0 protocol, patients were requested to elevate 1 leg at 45º for 5 seconds while supine. The SLR maneuver was considered effective when intra-abdominal pressure increased by 50%. IEPs were recorded 5 cm above the lower esophageal sphincter at baseline and during SLR. GERD was defined as AET greater than 6%.
RESULTS: The SLR was effective in 295 patients (81%), 115 (39%) of whom had an AET greater than 6%. Hiatal hernia (EGJ type 2 or 3) was seen in 135 (46%) patients. Compared with patients with an AET less than 6%, peak IEP during SLR was significantly higher in the GERD group (29.7 vs 13.9 mm Hg; P < .001). Using receiver operating characteristic analysis, an increase of 11 mm Hg of peak IEP from baseline during SLR was the optimal cut-off value to predict an AET greater than 6% (area under the receiver operating characteristic curve, 0.84; sensitivity, 79%; and specificity, 85%), regardless of the presence of hiatal hernia. On multivariable analysis, an IEP pressure increase during the SLR maneuver, EGJ contractile integral, EGJ subtype 2, and EGJ subtype 3, were found to be significant predictors of AET greater than 6% CONCLUSIONS: The SLR maneuver can predict abnormal an AET, thereby increasing the diagnostic value of HRM when GERD is suspected.
CLINICALTRIALS: gov ID: NCT04813029.
METHODS: Our aim was to determine normative EGJ metrics in a large international cohort of healthy volunteers undergoing HRM (Medtronic, Laborie, and Diversatek software) acquired from 16 countries in four world regions. EGJ-CI was calculated by the same two investigators using a distal contractile integral-like measurement across the EGJ for three respiratory cycles and corrected for respiration (mm Hg cm), using manufacturer-specific software tools. EGJ morphology was designated according to Chicago Classification v3.0. Median EGJ-CI values were calculated across age, genders, HRM systems, and regions.
RESULTS: Of 484 studies (28.0 years, 56.2% F, 60.7% Medtronic studies, 26.0% Laborie, and 13.2% Diversatek), EGJ morphology was type 1 in 97.1%. Median EGJ-CI was similar between Medtronic (37.0 mm Hg cm, IQR 23.6-53.7 mm Hg cm) and Diversatek (34.9 mm Hg cm, IQR 22.1-56.1 mm Hg cm, P = 0.87), but was significantly higher using Laborie equipment (56.5 mm Hg cm, IQR 35.0-75.3 mm Hg cm, P
MATERIALS AND METHODS: A review of observational studies was conducted to discuss the accuracy, tolerability and ease of use of tonometers in measuring IOP in children with glaucoma.
RESULTS: Goldmann applanation tonometry (GAT) and its portable handheld versions remain the gold standard in measuring IOP. Tono-Pen (Reichert Ophthalmic Instruments, Depew, New York, USA) and rebound tonometer (RBT) both correlate well with GAT. Although both tonometers tend to overestimate IOP, Tono-Pen overestimates more than RBT. Overestimation is more remarkable in higher IOP and corneal pathologies (such as but not limited to scarred cornea and denser corneal opacity). RBT was better tolerated than other tonometers in children and was easier to use in children of all ages.
CONCLUSIONS: RBT is the preferred tonometer for measuring IOP in children with glaucoma, as it is less traumatic, time efficient and does not require fluorescein dye or anaesthesia. However, examiners should use a second tonometer to confirm elevated IOP readings from the RBT.
METHOD: Eligible healthy Malay volunteers were invited to undergo the high-resolution esophageal manometry (inSIGHT Ultima, Diversatek Healthcare, Milwaukee, WI, USA). In recumbent and standing positions, test swallows were performed using liquid, viscous, and solid materials. Metrics including integrated relaxation pressure 4 s (IRP-4 s, mmHg), distal contractile integral (DCI, mmHg s cm), distal latency (DL, s), and peristaltic break (PB, cm) were reported in median and 95th percentile.
RESULTS: Fifty of 57 screened participants were recruited, and 586 saline, 265 viscous, and 261 solid swallows were analyzed. Per-patient wise, in the recumbent position, 95th percentile for IRP-4 s, DCI, DL, and PB were 16.5 mmHg, 2431 mmHg s cm, 8.5 s, and 7.2 cm, respectively. We observed that with each posture, the use of viscous swallows led to changes in DL, but the use of solid swallows led to more changes in the metrics including DCI and length of PB. Compared with a recumbent posture, anupright posture led to lower IRP-4 s and DCI values. Both per-patient analysis and per-swallow analyses yielded almost similar results when comparing the different postures and types of swallows. No major motility disorders were observed in this cohort of asymptomatic population. However, more motility disorders were reported in the upright position.
CONCLUSIONS: Variations in metrics can be observed in different postures and with different provocative swallow materials in a healthy population. The normative Chicago 3.0 metrics are also determined for the Malay population.
METHODS: We used the questionnaire previously established by our team for researchers in European countries. The correlation between the decreased rate of gastrointestinal motility and function tests, and the infection/mortality rates of COVID-19 and stringency of a government's interventions in each country was analysed and protective measures were assessed.
RESULTS: In total, 58 gastroenterologists/motility experts in Asian countries responded to this survey. The infection/mortality rates of COVID-19 and Stringency Index had a significant impact on the testing capacity of oesophageal manometry and catheter-based pH monitoring. In European countries, most facilities used filtering facepiece 2/3 (FFP2/3) masks during oesophageal motility studies. Meanwhile, in Asian countries, most facilities used surgical masks.
CONCLUSION: The total infection and mortality rates of COVID-19 can affect the rate of gastrointestinal motility testing and the type of protective equipment that must be used.