Abstract

Two important parameters for utilizing Ko-standardization method namely, absolute peak efficiency at reference position and peak-to-total ratio at different geometrical positions using standard point sources and HPGe were experimentally determined. Coincidence correction factor, C, for reference position and certain nuclides were also calculated and all almost equal to one. The importance and implication of this work to the K0-standardization method are presented. Other essential nuclear parameters which have to be experimentally determined or obtained from literature are also presented.
Dua parameter untuk kegunaan kaedah pemiawaian K0 iaitu kecekapan puncak absolut pada kedudukan rujukan dan nisbah jumlah-ke-puncak pada berbagai kedudukan geometri telah ditentukan secara eksperimen. Punca titik piawai dan HPGe digunakan. Faktor pembetulan koinsiden, C, yang digunakan untuk kedudukan piawai dan beberapa nuklid telah dihitung dan semuanya mempunyai nilai hampir sama dengan satu. Kepentingan dan implikasi kerja ini kepada kaedah pemiawaian K0 dikemukakan. Parameter nuklear lain yang ditentukan secara eksperimen atau yang diperolehi daripada pustaka juga dibentangkan.