METHODS: All relevant studies were identified through keyword searches in electronic databases from inception until September 2020. The searched publications were reviewed, categorised and analysed based on their respective methodology.
RESULTS: Hundred and one publications were identified which utilised existing MC-based applications/programs or customised MC simulations. Two outstanding challenges were identified that contribute to uncertainties in the virtual simulation reconstruction. The first challenge involves the use of anatomical models to represent individuals. Currently, phantom libraries best balance the needs of clinical practicality with those of specificity. However, mismatches of anatomical variations including body size and organ shape can create significant discrepancies in dose estimations. The second challenge is that the exact positioning of the patient relative to the beam is generally unknown. Most dose prediction models assume the patient is located centrally on the examination couch, which can lead to significant errors.
CONCLUSION: The continuing rise of computing power suggests a near future where MC methods become practical for routine clinical dosimetry. Dynamic, deformable phantoms help to improve patient specificity, but at present are only limited to adjustment of gross body volume. Dynamic internal organ displacement or reshaping is likely the next logical frontier. Image-based alignment is probably the most promising solution to enable this, but it must be automated to be clinically practical.
METHODS: Radiation dose received at left outer canthus (LOC) and left eyelid (LE) were measured using Metal-Oxide-Semiconductor Field-Effect Transistor dosimeters on 35 patients who underwent diagnostic or cerebral embolization procedures.
RESULTS: The radiation dose received at the LOC region was significantly higher than the dose received by the LE. The maximum eye lens dose of 1492 mGy was measured at LOC region for an AVM case, followed by 907 mGy for an aneurysm case and 665 mGy for a diagnostic angiography procedure. Strong correlations (shown as R(2)) were observed between kerma-area-product and measured eye doses (LOC: 0.78, LE: 0.68). Lateral and frontal air-kerma showed strong correlations with measured dose at LOC (AKL: 0.93, AKF: 0.78) and a weak correlation with measured dose at LE. A moderate correlation was observed between fluoroscopic time and dose measured at LE and LOC regions.
CONCLUSIONS: The MOSkin dose-monitoring system represents a new tool enabling real-time monitoring of eye lens dose during neuro-interventional procedures. This system can provide interventionalists with information needed to adjust the clinical procedure to control the patient's dose.
KEY POINTS: Real-time patient dose monitoring helps interventionalists to monitor doses. Strong correlation was observed between kerma-area-product and measured eye doses. Radiation dose at left outer canthus was higher than at left eyelid.
METHODS: We scrutinized the routine radiological exposure parameters during 58 clinical neuro-interventional procedures such as, exposure direction, magnification, frame rate, and distance between image receptor to patient's body and evaluate their effects on patient's dose using an anthropomorphic phantom. Radiation dose received by the occipital region, ears and eyes of the phantom were measured using MOSkin detectors.
RESULTS: DSA imaging technique is a major contributor to patient's dose (80.9%) even though they are used sparingly (5.3% of total frame number). The occipital region of the brain received high dose largely from the frontal tube constantly placed under couch (73.7% of the total KAP). When rotating the frontal tube away from under the couch, the radiation dose to the occipital reduced by 40%. The use of magnification modes could increase radiation dose by 94%. Changing the image receptor to the phantom surface distance from 10 to 40cm doubled the radiation dose received by the patient's skin at the occipital region.
CONCLUSION: Our findings provided important insights into the contribution of selected fluoroscopic exposure parameters and their impact on patient's dose during neuro-interventional radiology procedures. This study showed that the DSA imaging technique contributed to the highest patient's dose and judicial use of exposure parameters might assist interventional radiologists in effective skin and eye lens dose reduction for patients undergoing neuro-interventional procedures.
METHODS: 41 medical personnel performing 79 procedures were monitored for their eye lens exposure using the NanoDot™ optically-stimulated luminescence dosimeters (OSLD) taped to the outer canthus of their eyes. The air-kerma area product (KAP), fluoroscopy time (FT) and number of procedure runs were also recorded.
RESULTS: KAP, FT and number of runs were strongly correlated. However, only weak to moderate correlations were observed between these parameters with the measured eye lens doses. The average median equivalent eye lens dose was 0.052 mSv (ranging from 0.0155 to 0.672 mSv). The eye lens doses of primary operators were found to be significantly higher than their assistants due to the closer proximity to the patient and X-ray tube. The left eye lens of the operators received the highest amount of radiation due to their habitual positioning towards the radiation source.
CONCLUSION: KAP and FT were not useful in predicting the equivalent eye lens dose exposure in interventional radiological procedures. Direct in vivo measurements were needed to provide a better estimate of the eye lens doses received by medical personnel during these procedures. This study highlights the importance of using direct measurement, such as OSLDs, instead of just indirect factors to monitor dose in the eye lens in radiological procedures.