MATERIAL AND METHOD: Forty sound permanent mandibular premolars were collected from a dental clinical setting and disinfected. All forty samples were mounted vertically in a rubber mold exposing only the clinical crown. All samples were bleached using Opalescence Boost Professional Teeth Whitening. After the bleaching procedure, each sample was randomly allocated into four groups according to surface treatment. Samples in group 1 were treated with methylene blue photosensitizer (MBP). Samples in group 2 were exposed to 10% sodium ascorbate. Samples in group 3 were treated with Er, Cr: YSGG laser (ECL). Samples in group 4 were not treated (control). All Samples were treated with 37% phosphoric acid and a bonding agent was applied. A bulk-fill composite was cured to all specimens and all samples were treated in a thermocycler. Specimens were placed in a universal testing machine for shear bond strength (SBS) testing. Descriptive statistics were associated by analysis of variance (ANOVA) and Tukey's post hoc test maintaining level of significance (p<0.05) RESULTS: The lowest SBS scores were achieved in the bleached enamel (BE) group (15.25±1.745 MPa). Whereas, the highest bond integrity was attained by AA group (32.23±1.854 MPa). Samples treated with ECL (31.87±1.659 MPa) and AA (32.23±1.854) were comparable (p>0.05). Samples treated with PDT exhibited significantly different SBS (22.41±1.258) compared to other experimental groups CONCLUSION: ECL showed a reversal effect of BE compared to AA and has the potential to be used in clinical settings. BE reversal using MBP needs further investigation.
OBJECTIVE: The purpose of this study was to evaluate the system's performance during treadmill tests to maximum exertion in a subset of patients within the Micra Transcatheter Pacing Study.
METHODS: Patients underwent treadmill testing at 3 or 6 months postimplant with algorithm programming at physician discretion. Normalized sensor rate (SenR) relative to the programmed upper sensor rate was modeled as a function of normalized workload in metabolic equivalents (METS) relative to maximum METS achieved during the test. A normalized METS and SenR were determined at the end of each 1-minute treadmill stage. The proportionality of SenR to workload was evaluated by comparing the slope from this relationship to the prospectively defined tolerance margin (0.65-1.35).
RESULTS: A total of 69 treadmill tests were attempted by 42 patients at 3 and 6 months postimplant. Thirty tests from 20 patients who completed ≥4 stages with an average slope of 0.86 (90% confidence interval 0.77-0.96) confirmed proportionality to workload. On an individual test basis, 25 of 30 point estimates (83.3%) had a normalized slope within the defined tolerance range (range 0.46-1.08).
CONCLUSION: Accelerometer-based rate adaptive pacing was proportional to workload, thus confirming rate adaptive pacing commensurate to workload is achievable with an entirely intracardiac pacing system.