FINDINGS: A total of 274 venous blood was collected from normal healthy adults during the community screening programmes. The performance of POC devices, Afinion and Quo-test were compared to central laboratory HPLC method; Adams A1c HA 8160. Both POC devices showed good correlation to HA 8160 with r = 0.94 (p < 0.001) and r = 0.95 (p < 0.001) for Afinion and Quo-test respectively. The means difference were statistically higher between POC and HA 8160 with 0.23% (95% CI 0.19-0.26, p < 0.001) and 0.29% (95% CI 0.24-0.34, p < 0.001) for Afinion and Quo-test respectively.
CONCLUSIONS: Both POC devices could be considered in health clinics for diabetes management but not to be used for the diagnostic purposes.