Material and Methods: Four FNEs were retrieved from revision surgeries of four patients with prior intramedullary nail fixation of their pertrochanteric hip fractures complicated by femoral head perforation. The FNEs were divided into two groups based on whether or not there was radiographic evidence of medial migration prior to the revisions. Wear patterns on the FNEs were then assessed using both scanning electron microscopy and light microscopy.
Results: Repetitive, linearly-arranged, regularly-spaced, unique transverse scratch marks were found only in the group with medial migration, corresponding to the specific segment of the FNE that passed through the intramedullary component of the PFNA during medial migration. These scratch marks were absent in the group without medial migration.
Conclusion: Our findings are in support of a ratcheting mechanism behind the medial migration phenomenon with repetitive toggling at the intramedullary nail-FNE interface and progressive propagation of the FNE against gravity.