PIP: In 1986, single puncture laparoscopic application of the Filshie clip was introduced into Malaysia's National Population and Family Development Board's female sterilization program for an evaluation. A total of 42 female sterilizations were performed by this method between June 1986 and December 1986. Laparoscopy was performed on an outpatient basis under local anesthesia. In 37 cases, the sterilization was interval. Filshie clips were successfully applied to the fallopian tubes of 40 women; in the remaining 2 cases, the latching mechanism jammed and Falope rings were used instead. The average time from application of 1st clip to skin closure was 10 minutes. During clip application, technical problems (faulty applicator, clip fell into the pelvic cavity, failure of latching mechanism) were encountered in 4 cases and surgical difficulties (poor visualization, hydrosalpinx, and tubal adhesions) occurred in 10 cases. In general, complications were minimal: 1 patient developed cynosis after anesyhesia (Pethidine) injection, the omentum was torn during introduction of the trocar in another, and a 3rd patient developed thrombophlebitis at the intravenous site. Only 1 pregnancy was recorded during the follow-up period; this occurred 15 months postoperatively and was intrauterine. Given the facts that this procedure is easy to learn, fast and simple, associated with no more complications than other laparoscopic procedures, and is cosmetically more acceptable to women, its more widespread application is recommended.