METHODS: Fifty-one participants performed the standard incremental treadmill exercise in a controlled laboratory setting with 12-lead ECG attached to the patient's body and wearing wrist-worn PPG trackers.
RESULTS: At each stage, the absolute percentage error of the PPG was within 10% of the standard acceptable range. Further analysis using a linear mixed model, which accounts for individual variations, revealed that PPG yielded the best performance at the baseline low-intensity exercise. As the stages progressed, heart rate validity decreased but was regained during recovery. The reliability was moderate to excellent.
CONCLUSIONS: Low-cost trackers AMAZFIT Cor and Bip validity and reliability were within acceptable ranges, especially during low-intensity exercise among patients with ischemic heart disease recovering from cardiac procedures. Though using the tracker as part of the diagnosis tool still requires more supporting studies, it can potentially be used as a self-monitoring tool with precautions.