Abstract

The aimed of this study is to estimate the entrance surface dose (ESD) of routine chest X-ray (CXR) examination and to compare the direct and indirect measurement. This study also extended to relate with the patients body thickness and established lifetime attributable risk (LAR) of the examinations. A total of 101 CXR examinations in posteroanterior (PA) projections were selected
as subjects and data, such as patient habitus and scanning acquisition parameters were recorded. The mean ESD value obtained from the TLD-100 and the software calculation was 0.31 mGy and 0.17 mGy, respectively. The percentage deviation obtained ranged from 25.5% to 61.3%. In comparison with the national diagnostic reference level (DRL), the mean values of ESD obtained from this study were lower by a factor of 3. Furthermore, it was observed that the mean absorbed dose of the adrenals, kidneys, lungs, oesophagus, and heart walls were lower as compared to the ESD value. The examinations were only associated with very low risk of cancer incidence and mortality. Hence, the results of this study suggested a need for standardisations of the personnel training to perform X-ray examinations according to ALARA principles.