Int Ophthalmol, 2015 Aug;35(4):459-66.
PMID: 25024102 DOI: 10.1007/s10792-014-9970-4

Abstract

To determine the accuracy of intraocular lens (IOL) calculations in eyes undergoing phacoemulsification cataract surgery with IOL implantation using immersion A-scan ultrasound (US) and Lenstar LS 900(®) biometry. In this prospective study, 200 eyes of 200 patients were randomized to undergo either Lenstar LS 900(®) or immersion A-scan US biometry to determine the IOL dioptric power prior to phacoemulsification cataract surgery. Post-operative refractive outcomes of these two groups of patients were compared. The result showed no significant difference between the target spherical equivalent (SE) and the post-operative SE value by the Lenstar LS 900(®) (p value = 0.632) or immersion A-scan US biometry (p value = 0.438) devices. The magnitude of difference between the two biometric devices were not significantly different (p value = 0.868). There was no significant difference in the predicted post-operative refractive outcome between immersion A-scan US biometry and Lenstar LS 900(®). Based on the results, the immersion A-scan US technique is as accurate as Lenstar LS 900(®) in the hands of an experienced operator.