METHODS: In the present study, 2D axisymmetric models were developed to investigate how saline backflow influence saline-infused RFA and whether the aforementioned concerns are warranted. Saline-infused RFA was described using the dual porosity-Joule heating model. The hydrodynamics of backflow was described using Poiseuille law by assuming the flow to be similar to that in a thin annulus. Backflow lengths of 3, 4.5, 6 and 9 cm were considered.
RESULTS: Results showed that there is no concern of thermally ablating the tissue in the backflow region. This is due to the Joule heating being inversely proportional to distance from the electrode to the fourth power. Results also indicated that larger backflow lengths led to larger growth of thermal damage along the backflow region and greater decrease in coagulation volume. Hence, backflow needs to be controlled to ensure an effective treatment of saline-infused RFA.
CONCLUSIONS: There is no risk of ablating tissues around the needle insertion track due to backflow. Instead, the risk of underablation as a result of the loss of saline due to backflow was found to be of greater concern.
METHODS: Radiofrequency and microwave ablation of liver tumours were performed on 20 patients (40 lesions) with the assistance of a CT-guided robotic positioning system. The accuracy of probe placement, number of readjustments and total radiation dose to each patient were recorded. The performance level was evaluated on a five-point scale (5-1: excellent-poor). The radiation doses were compared against 30 patients with 48 lesions (control) treated without robotic assistance.
RESULTS: Thermal ablation was successfully completed in 20 patients with 40 lesions and confirmed on multiphasic contrast-enhanced CT. No procedure related complications were noted in this study. The average number of needle readjustment was 0.8 ± 0.8. The total CT dose (DLP) for the entire robotic assisted thermal ablation was 1382 ± 536 mGy.cm, while the CT fluoroscopic dose (DLP) per lesion was 352 ± 228 mGy.cm. There was no statistically significant (p > 0.05) dose reduction found between the robotic-assisted versus the conventional method.
CONCLUSION: This study revealed that robotic-assisted planning and needle placement appears to be safe, with high accuracy and a comparable radiation dose to patients.
KEY POINTS: • Clinical experience on liver thermal ablation using CT-guided robotic system is reported. • The technical success, radiation dose, safety and performance level were assessed. • Thermal ablations were successfully performed, with an average performance score of 4.4/5.0. • Robotic-assisted ablation can potentially increase capabilities of less skilled interventional radiologists. • Cost-effectiveness needs to be proven in further studies.
METHODS: We report our preliminary experience of performing radiofrequency ablation of the liver using a robotic-assisted CT guidance system on 11 patients (17 lesions).
RESULTS/CONCLUSION: Robotic-assisted planning and needle placement appears to have high accuracy, is technically easier than the non-robotic-assisted procedure, and involves a significantly lower radiation dose to both patient and support staff.
KEY POINTS: • An early experience of robotic-assisted radiofrequency ablation is reported • Robotic-assisted RFA improves accuracy of hepatic lesion targeting • Robotic-assisted RFA makes the procedure technically easier with significant lower radiation dose.
METHODS: CFAE from several atrial sites, recorded for a duration of 16 s, were acquired from 10 patients with persistent and 9 patients with paroxysmal AF. These signals were appraised using non-overlapping windows of 1-, 2- and 4-s durations. The resulting data sets were analyzed with Recurrence Plots (RP) and Recurrence Quantification Analysis (RQA). The data was also quantified via entropy measures.
RESULTS: RQA exhibited unique plots for persistent versus paroxysmal AF. Similar patterns were observed to be repeated throughout the RPs. Trends were consistent for signal segments of 1 and 2 s as well as 4 s in duration. This was suggestive that the underlying signal generation process is also repetitive, and that repetitiveness can be detected even in 1-s sequences. The results also showed that most entropy metrics exhibited higher measurement values (closer to equilibrium) for persistent AF data. It was also found that Determinism (DET), Trapping Time (TT), and Modified Multiscale Entropy (MMSE), extracted from signals that were acquired from locations at the posterior atrial free wall, are highly discriminative of persistent versus paroxysmal AF data.
CONCLUSIONS: Short data sequences are sufficient to provide information to discern persistent versus paroxysmal AF data with a significant difference, and can be useful to detect repeating patterns of atrial activation.