METHODS: The 3D-printed cardiac insert phantom was positioned into a chest phantom and scanned with a 16-slice CT scanner. Acquisitions were performed with CCTA protocols using 120 kVp at four different tube currents, 300, 200, 100 and 50 mA (protocols A, B, C and D, respectively). The image data sets were reconstructed with a filtered back projection (FBP) and three different IR algorithm strengths. The image quality metrics of image noise, signal-noise ratio (SNR) and contrast-noise ratio (CNR) were calculated for each protocol.
RESULTS: Decrease in dose levels has significantly increased the image noise, compared to FBP of protocol A (P
METHODOLOGY: This prospective, randomized, controlled trial enrolled pregnant females undergoing LSCS after spinal anesthesia. The participants were randomized into two groups: a novel 3D device group and a traditional wedge group. Primary outcome measures included the incidence of SHS, while secondary outcomes included maternal hemodynamic parameters, fetal outcomes, feasibility, ease of use, and the safety profile of the devices.
RESULTS: Baseline characteristics were well balanced between the two groups. Although some differences in maternal hemodynamic parameters were noted, the incidence of SHS was significantly lower in the novel 3D device group than that in the traditional wedge group. Fetal outcomes did not differ significantly between the groups. The novel 3D device demonstrated high compatibility with various patient anatomies and was easy to integrate into routine practice. The adverse event profiles were similar between the groups.
CONCLUSION: This study highlights the potential of a novel 3D-printed uterine displacement device for preventing SHS during LSCS, thereby improving maternal and fetal outcomes. Future research should further validate these findings and explore the long-term implications of the maternal and neonatal outcomes.
METHODS: A prospective study was conducted to evaluate the outcome of a new reconstruction technique that produces patient-specific hybrid polymethyl methacrylate-autologous cranial implant. Computer-assisted 3D modeling and printing was utilized to produce patient-specific molds, which allowed real-time reconstruction of bone flap with partial defect intra-operatively.
RESULTS: Outcome assessment for 11 patients at 6 weeks and 3 months post-operatively revealed satisfactory implant alignment with favorable cosmesis. The mean visual analog scale for cosmesis was 91. Mean implant size was 50cm, and the mean duration of intra-operative reconstruction was 30 minutes. All of them revealed improvement in quality of life following surgery as measured by the SF-36 score. Cost analysis revealed that this technique is more cost-effective compared to customized cranial prosthesis.
CONCLUSION: This new technique and approach produce hybrid autologous-alloplastic bone flap that resulted in satisfactory implant alignment and favorable cosmetic outcome with relatively low costs.
Methods: A patient-specific 3D-printed breast model was generated using 3D-printing techniques for the construction of the hollow skin and fibroglandular region shells. Then, the T1 relaxation times of the five selected materials (agarose gel, silicone rubber with/without fish oil, silicone oil, and peanut oil) were measured on a 3T MRI system to determine the appropriate ones to represent the MR imaging characteristics of fibroglandular and adipose tissues. Results were then compared to the reference values of T1 relaxation times of the corresponding tissues: 1,324.42±167.63 and 449.27±26.09 ms, respectively. Finally, the materials that matched the T1 relaxation times of the respective tissues were used to fill the 3D-printed hollow breast shells.
Results: The silicone and peanut oils were found to closely resemble the T1 relaxation times and imaging characteristics of these two tissues, which are 1,515.8±105.5 and 405.4±15.1 ms, respectively. The agarose gel with different concentrations, ranging from 0.5 to 2.5 wt%, was found to have the longest T1 relaxation times.
Conclusions: A patient-specific 3D-printed breast phantom was successfully designed and constructed using silicone and peanut oils to simulate the MR imaging characteristics of fibroglandular and adipose tissues. The phantom can be used to investigate different MR breast imaging protocols for the quantitative assessment of breast density.
OBJECTIVES: The aim of the study was to evaluate the deviation of implant placement performed with a surgical guide fabricated by means of the rapid prototyping technique (the PolyJet™ technology).
MATERIAL AND METHODS: Twenty sheep mandibles were used in the study. Pre-surgical cone-beam computed tomography (CBCT) scans were acquired for the mandibles by using the Kodak 9000 3D cone-beam system. Two implants with dimensions of 4 mm in diameter and 10 mm in length were virtually planned on the 3D models of each mandible by using the Mimics software, v. 16.0. Twenty surgical guides were designed and printed using the PolyJet technology. A total of 40 implants were placed using the surgical guides, 1 on each side of the mandible (2 implants per mandible). The post-surgical CBCT scans of the mandibles were performed and superimposed on the pre-surgical CBCT scans. The amount of deviation between the virtually planned placement and the actual implant placement was measured, and a descriptive analysis was done.
RESULTS: The results showed that the mean deviation at the implant coronal position was 1.82 ±0.74 mm, the mean deviation at the implant apex was 1.54 ±0.88 mm, the mean depth deviation was 0.44 ±0.32 mm, and the mean angular deviation was 3.01 ±1.98°.
CONCLUSIONS: The deviation of dental implant placement performed with a 3D-printed surgical guide (the PolyJet technology) is within the acceptable 2-millimeter limit reported in the literature.