CASE REPORT: This paper describes a case of SOT affecting the anterior mandible of a 10-year-old Indian female. The patient was treated by local surgical excision and there has been no follow-up clinical record of recurrence 5 years after primary treatment. Histo?pathological examination revealed a solid, locally-infiltrative neoplasm composed of bland-looking squamatoid islands scattered in a mature fibrous connective tissue stroma and the diagnosis was SOT. Immunohistochemical evaluation showed positive reactivity of varying intensity in the neoplastic epithelial cells for Notch1, Notch3, Notch4, and their ligands Jagged1 and Delta1. Expression patterns showed considerable overlap. No immunoreactivity was detected for Notch2 and Jagged2.
CONCLUSIONS: Present findings suggest that Notch receptors and their ligands play differential roles in the cytodifferentiation of SOT.
METHODS: A 160-mm-diameter PLA head CTDI phantom with five dosimetry holes was fabricated using fused deposition modeling. Its elemental composition, density, effective atomic number, electron density, Hounsfield units (HU), and CT dosimetry parameters were characterized. Measurements were performed on 128-slice Ingenuity Philips and GE Discovery scanners, at 80, 120, and 140 kV. CTDI was measured at two different times and then compared with PMMA.
RESULTS: PLA showed properties close to PMMA, with differences of 5.08% in density, -0.015% in effective atomic number, and -4.46% in electron density. PLA HU values were lower than PMMA, ranging from 22.5 to 45.3 HU on Philips and 25.1-49.1 HU on GE. CTDIvol deviations between PLA and PMMA ranged from -9.00% to -2.82% in November 2025 and from -5.78% to -3.83% in June 2026. These deviations remained within <10%, which was generally considered acceptable for comparative phantom-based CT dosimetry.
CONCLUSION: The PLA CTDI phantom showed good effective atomic number, electron density, and dosimetric agreement with the standard PMMA phantom across two CT systems and two measurement time points.
IMPLICATIONS FOR PRACTICE: 3D-printed PLA CTDI phantom could offer a practical, low-cost complementary tool for comparative CTDI measurement, education, protocol optimization, and quality assurance.