METHODS: Magnetic resonance imaging scans of the head of a patient with a large left frontal cyst were segmented to construct a finite element model to study the E-field distribution during ECT. Five electrode configurations were investigated: right unilateral, left unilateral, bifrontal, and bitemporal and left anterior right temporal. The E-field distributions for all montages were compared with a hypothetical condition where brain tissue and electrical conductivity from the right frontal region was mirrored across the longitudinal fissure into the cyst.
RESULTS: Differences in mean E-field and 90th percentile E-fields were mainly observed in brain regions closest to the cyst including the left inferior frontal gyrus and left middle frontal gyrus. This trend was most pronounced in montages where the electrodes were closest to the cyst such as left unilateral and bitemporal.
CONCLUSION: The presence of a highly conductive cyst close to the ECT electrode tended to attract current into the cyst region, altering current pathways, with potential implications for therapeutic efficacy and safety. Placing electrodes farther away from the cyst is likely to minimize any effects on the E-field distribution and potentially clinical outcomes.