MATERIALS AND METHODS: Seventy-four participants, with type 1 (T1D, n = 24), type 2 (T2D, n = 11), or gestational (n = 39) diabetes, were enrolled across 13 sites (9 in United Kingdom, 4 in Austria). Average gestation was 26.6 ± 6.8 weeks (mean ± standard deviation), age was 30.5 ± 5.1 years, diabetes duration was 13.1 ± 7.3 years for T1D and 3.2 ± 2.5 years for T2D, and 49/74 (66.2%) used insulin to manage their diabetes. Sensors were worn for up to 14 days. Sensor glucose values (masked) were compared with capillary SMBG values (made at least 4 times/day).
RESULTS: Clinical accuracy of sensor results versus SMBG results was demonstrated, with 88.1% and 99.8% of results within Zone A and Zones A and B of the Consensus Error Grid, respectively. Overall mean absolute relative difference was 11.8%. Sensor accuracy was unaffected by the type of diabetes, the stage of pregnancy, whether insulin was used, age or body mass index. User questionnaires indicated high levels of satisfaction with sensor wear, system use, and comparison to SMBG. There were no unanticipated device-related adverse events.
CONCLUSIONS: Good agreement was demonstrated between the FreeStyle Libre System and SMBG. Accuracy of the system was unaffected by patient characteristics, indicating that the system is safe and accurate to use by pregnant women with diabetes.